Code
Name
0800
File open error
0801
File format error
0802
File setup value error
0803
ID information read error
0804
Image file close error
0805
Image file read error
0806
Format information search error
0807
Format number acquisition error
0809
DMA device initialization error
080A
DMA transfer error
080B
Failure to reserve memory area
0810
DSP10 device initialization error
0811
DSP10 µ load error
0812
DSP10 memory write error
009-051-04
05.31.99 FM2460
5
6
Significance/Occurrence Condition
The HD is faulty or the file is not registered.
The file structure is improper.
The data value of the file is improper.
Faulty HD
Faulty HD
Faulty HD
• Format information file in the hard disk (the data
in LPFMT0J.DAT/LPFMT1J.DAT is improper).
• The format information for the battery backup memory
is improper.
The combination of read size and format is improper.
The DMA device cannot be opened.
DMA image transfer cannot be done.
Insufficient memory area
The DSP device cannot be opened.
The DSP µ is not yet registered in the HD.
Write to the DSP device cannot be done.
CR-IR341 Service Manual
Detail: Denotes the detail information format.
Probable Cause/Remedy
Replace the HD or register the file.
Register the file again.
Register the file again.
Replace the HD or reboot.
Replace the HD or reboot.
Replace the HD or reboot.
• Reinstall the software.
• Clear the battery backup memory.
Check the read size, etc.
Check the IMG08A board.
Check the IMG08A board.
Reboot.
Check the IMG08A board.
Register the DSP µ in the HD again.
Check the IMG08A board.
MT - 13
Detail
Refer to
I-1
I-1
I-1
I-2
I-2
I-2
I-3
I-4
I-5
I-5
I-6
I-7
I-7
I-7