FujiFilm CR-IR341 Service Manual page 307

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
3
4
ONTENTS
Code
Name
2A40
Barcode data error
2B18
DSP H image processing error
2B30
Compressed/decompressed image left
2C40
Unexpected command received
2C41
Image transfer error
2C42
Switching IDT connection to CPU90E
2CC0
Unexpected command received
2CD0
20 line/mm image acceptance
2D00
SCU-related information is illegal
2D02
SCP-related information is illegal
2D03
PRINT output not allowed
2D04
Soft key invalid
2D05
SOP class designation is illegal
2D06
Older version of HCP board
009-051-07
08.30.2001 FM3115
5
6
Significance/Occurrence Condition
The data read by the barcode reader is improper
(without STX or EXT code).
An error occurs during compression/decompression
processing.
Although image data is left to be compressed or
decompressed, the compression/decompression is
completed.
An unexpected command is received.
An error occurs during image data transfer.
Because the CPU90F board does not boot normally, the
connection to the IDT is switched from CPU90F to CPU90E
in the configuration setting.
An unexpected command is received.
Because the machine cannot handle 20 line/mm images,
input is rejected.
There is no "Base on DICOM" information for the own host
(DICOM Storage Service SCU).
There is no "Base on DICOM" information corresponding to
the output designation host (DICOM Storage Service SCP).
The output destination attribute in the device information is
PRINT only.
There is no "Base on DICOM" information for the own host
(DICOM Storage Service SCU).
The "Base on DICOM" soft key is inconsistent with the SOP
class designation for the output designation host in the
"Base on DICOM" information file.
The HCP board is not a version that supports 11bit output
(QA-WS output is 11bit).
CR-IR341 Service Manual
Detail: Denotes the detail information format.
Probable Cause/Remedy
Barcode reader, SNS08A board
Faulty HDD, MTH board, HCP board, etc.
The image data is corrupted.
Check the sending/receiving status on the target end.
Check the Ethernet cable; the Ethernet cable is faulty;
or check the sending/receiving status on the target end.
• The CPU90F board does not boot normally.
• The communication between the CPU90E board and
CPU90F board (DMA/2-port memory) is faulty.
Check the sending/receiving status on the unit
connected.
• No host file
• Base on DICOM information definition is incorrect.
• HD failure
• No host file
• Base on DICOM information definition is incorrect.
• HD failure
An attempt is made to provide output to a device that
only supports the PRINT service.
• No host file
• Base on DICOM information definition is incorrect.
• HD failure
• Base on DICOM information definition is incorrect.
• Base on DICOM soft key has been installed
improperly.
Replace the HCP board with its version G or later.
MT - 44
Detail
Refer to
K-2
L-2
L-1
M-3
M-5
D-2
M-3
G-1
G-1
G-1
F
G-1
F
TR1H1259.EPS
MT - 44

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir341pFcr 5000Fcr 5000plus

Table of Contents