FujiFilm CR-IR341 Service Manual page 501

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
To Enable Sensor Checks by Performing MECHANICAL UTILITY Operation
There are two methods for monitoring sensors: one is to enable monitoring of all sensors,
and the other is to select an individual sensor and enable its monitoring.
To enable monitoring of all sensors
(1) Select "6. MECHANICAL UTILITY" and "3. SENSOR" in sequence.
(2) Select "3. MONITOR ALL".
A list of sensor numbers is displayed so that the machine is placed in a sensor
monitoring enabled state.
0. QUIT
1. ERROR LOG UTILITY
2. CONFIGURATION SETTING
3. TEST MODE
4. ELECTRICAL UTILITY
5. SCANNER UTILITY
6. MECHANICAL UTILITY
7. FILE UTILITY
8. BACKUP MEMORY
9. HV OFF
10. MENU SETTING
> 6
0.QUIT
1.MOTOR
2.ACTUATOR
3.SENSOR
4.UNIT
MU > 3
0.QUIT
1.NUMBER
2.MONITOR
3.MONITOR ALL
MU:SEN > 3
MU:SEN > 3
SA1-17
SB1-5
SC1-2
SD1-3
SE1
SF1-4
SZ2-3
0:QUIT
SA1 ----> Close
SA2 ----> Open
SA2 ----> Close
SA1 ----> Open
0
0.QUIT
1.NUMBER
2.MONITOR
3.MONITOR ALL
MU:SEN >
009-051-04
05.31.99 FM2460
3
4
5
6
: 0 0 0 1 0 0 0 1 0 0 0 1 0 0 0 1 1
: 1 0 1 0 0
: 0 0
: 1 1 0
: 1
: 0 0 0 0
: 1 0
CR-IR341 Service Manual
7
8
9
10
A
B
C
D
E
F
7
8
9
4
5
6
1
2
3
0
.
SP
DEL
BS
ENT
Caps
A
B
C
D
E
F
7
8
9
4
5
6
1
2
3
0
.
SP
DEL
BS
ENT
Caps
FR1H1514.EPS
MC - 10