FujiFilm CR-IR341 Service Manual page 309

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
3
4
ONTENTS
Code
Name
3728
Network reception error
3729
ID delete response wait timeout
372A
File transfer
372B
Network transmission error
372C
Network transmission error
372D
Network transmission error
3751
File setup value error
3800
Receive line break detection
3801
LP status error
3802
LP level 0 error
3803
LP error recovery failure
3804
Character string transmission error
3805
No FINP ID
3806
Board insertion error
009-051-07
08.30.2001 FM3115
5
6
Significance/Occurrence Condition
Data reception over the network fails.
While waiting for an ID delete response to the network IDT,
a timeout occurs.
An ID acquisition request transmission to the network IDT
fails.
An ID delete request transmission to the network IDT fails.
A utility information transmission to the network IDT fails.
A file setup value check error (siosetup.idt) occurs.
A receive line break occurs.
An LP line status error is detected.
Level 0 error (system down) for the LP is detected.
Recover from level 1 error of the LP fails.
When a character string is transmitted, an end-of-
transmission error occurs.
The ID information (Finp) for the designating tag
(group/element) does not exist.
The IMG08M board (for 5000MA) is inserted inadvertently
into the machine.
CR-IR341 Service Manual
Detail: Denotes the detail information format.
Probable Cause/Remedy
The network environment is faulty.
The network environment is faulty.
The network environment is faulty.
The network environment is faulty.
The network environment is faulty.
The parameter value for the data-link layer parameter
file is out of range.
Check the line.
Check the line.
Check the LP.
Check the LP.
Check the LP.
Check the ID information.
Board insertion mistake
MT - 46
Detail
Refer to
H
H
H
H
H
A-1
I-6
I-6
I-6
I-6
I-8
I-12
TR1H1261.EPS
MT - 46