14. 9. HV ON/OFF: High-Voltage Switch (Software
Switch)
It should be turned OFF before opening the covers for checking. If not, the photomultiplier
may be damaged.
For the high-voltage switch, "9. HV OFF" is displayed when menu number "9" is in ON
status, while "9. HV ON" is displayed when it is in OFF status.
To turn high-voltage switch OFF
When the high-voltage switch is in the ON position, menu item 9 is displayed as "9.HV
OFF".
0.QUIT
1.ERROR LOG UTILITY
2.CONFIGURATION SETTING
3.TEST MODE
4.ELECTRICAL UTILITY
5.SCANNER UTILITY
6.MECHANICAL UTILITY
7.FILE UTILITY
8.BACKUP MEMORY
9.HV OFF
10.IP UTILITY
>
(1) Select "9. HV OFF".
The high-voltage switch is set to OFF, with the following message displayed.
THE HV IS TURNED OFF.
(2) Select "0. QUIT".
009-051-05