FujiFilm CR-IR341 Service Manual page 134

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
2.5
Feed Conveyance (IP Length Measurement)
(1) Overall/General Sequence
In this sequence, the IP sent from the IP removal unit is fed through the erasure conveyor to
the side-positioning conveyor.
During the conveyance, IP length measurement is performed.
If the IP length measurement is completed normally, the erasure lamp, if unlit, is lit.
Fig. 2.5.1 Passage through Before-Side-Positioning Conveyor and IP Length Measurement
(Overall Sequence)
(Sequence initiation requirement)
When the following requirement is met, the
sequence starts.
¥ IP feed conveyance permission OK: Fig. 2.9.1
To the conveyance path of the up-down IP
removal unit, the IP is not fed with the driving
motor, but with the belt drive, from the
before-side-positioning conveyor.
Data for determination of IP type is obtained.
Table 2.5.1 Operation Parameters
Symbol
PB22
Conveyance (feed)
PC11
Conveyance (feed)
PD31
Conveyance (feed)
Relationship between Conveyance Path
and IP Driving
SF4
SF2
SF1
009-051-04
05.31.99 FM2460
3
4
5
6
The erasure
lamp is lit.
Description
Value
+ infinite P
- infinite P
- infinite P
TR1H0021.EPS
SB2
IP
SF3
SC2
SC1
BCR
CR-IR341 Service Manual
7
8
9
10
Preprocessing: Fig. 2.1.1, and Fig. 3.1.1
START
IP feed conveyance
permitted?
N
Y
Conveyance driving
DRICE (MD3/PD31)
DRIVE (MC1/PB22)
DRIVE (MB2/PB22)
IP length measurement
Fig.2.5.2
Erasure lamp lit?
Y
N
Erasure lamp lit
DRIVE (LAMP 1-5/ON)
END
Structure and Operation of Path Changeover Mechanism
The conveyance path of the before-side-positioning
conveyor is normally in the feed state.
Upon load request, the convergence path changeover
guide driving solenoid SOLC1 is turned ON to implement
path changeover.
After the IP has been passed (sensor SB2 is CLOSE
the solenoid is turned OFF to reset it to its original state.
Guide
when SOLC1
is OFF
SOLC1
SC2
IP path
Before-side-positioning
conveyor
MD - 71
IP determination
Fig.2.5.3
FR1H0326.EPS
OPEN),
Guide
when SOLC1
is ON
SC1
IP path
FR1H0328.EPS
MD - 71