FujiFilm CR-IR341 Service Manual page 769

Fuji computed radiography
Table of Contents

Advertisement

C
11
12
ONTENTS
16.3 CPU90E Board
CAUTIONS
• After the CPU90E board is replaced, clear the contents of the backup memory and set the
date and time.
"2.3 Clearing Backup Memory, 2.4 Setting Date and Time"
• When servicing any printed circuit board, be sure to wear an anti-static wristband to ensure
proper grounding. If not, static electricity built on the human body may cause damage to
electric parts mounted on the board.
• If the lithium cell is drained, replace the board as a whole.
• The old board replaced should be sent back to the Parts Center (factory).
• If the cell is replaced improperly, it may explode. When the lithium cell is to be replaced, a cell
designated by Fuji or its equivalent should be used. The drained cell should be disposed of in
accordance with the instructions provided by Fuji.
Setting CPU90E Board
S1
OFF position
8
7
6
ON position
CN2
S5
STRAIGHT
position
009-051-04
05.31.99 FM2460
13
14
15
16
S7
OFF position
5
4
3
2
1
4
ON position
NFB4
NFB2
S6
CROSS
RS-422
position
position
CR-IR341 Service Manual
S2
OFF position
3
2
1
4
3
2
ON position
CN1
NFB3
NFB1
RS-232C
position
FR1B1149.EPS
MC - 274
1
MC - 274

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir341pFcr 5000Fcr 5000plus

Table of Contents