Cassette Set Unit - FujiFilm CR-IR341 Service Manual

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
(2) Insert a 300 mm scale along the right-hand side of the lowermost shelf (fourth shelf).
Press the manual release lever to lower the hold pin.
(3) Slide the multi-stage cassette loading cover until the scale comes into contact with the
cassette set unit entry reference guide protrusion (level difference: 0.5 – 1 mm), and
then tighten the four screws that were loosely retained.
(2)
Scale
Lower most shelf
009-051-04
009-051-06
04.20.2001 FM3006
05.31.99 FM2460
3
4
5
6
CR-IR341 Service Manual
7
8
9
10
(2)
Manual release
lever
Top view
(3)

Cassette set unit

entry reference guide
0.5 - 1mm
(white plastic)
Scale
MC - 24
0.5 - 1mm
Multi-stage
cassette loading cover
FR1H1485.EPS
MC - 24