FujiFilm CR-IR341 Service Manual page 221

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
6.4
Error Handling: Overexposure
(1) Overall/General Sequence
This error handling is performed when overexposed IP is determined in the erasure check
sequence.
Corresponding sequence
Fig. 2.10.5 Erasure Check
Fig. 6.4.1 Error Handling: Overexposure
At the time when the
error IP is returned to
the cassette, an elevation
permission is issued (via
normal routine) to start the
processing for the next IP
in the queue. At the same
time, removal of the cassette
for the error IP is notified to
the user.
6.5
Error Handling: Lamp Failure
(1) Overall/General Sequence
This error handling is performed when an unlit erasure lamp is detected in the erasure
check sequence.
Corresponding sequence
Fig. 2.10.5 Erasure Check
Fig. 6.5.1 Error Handling: Lamp Failure
At the time when the
error IP is returned to
the cassette, an elevation
permission is issued (via
normal routine) to start the
processing for the next IP
in the queue. At the same
time, removal of the cassette
for the error IP is notified to
the user.
009-051-04
05.31.99 FM2460
3
4
5
6
Preprocessing: Fig. 2.10.5
START
Error message output
Stop ARARM/UNLOAD CASS.
Alarm disabled
Cassette removed
END
Preprocessing: Fig. 2.10.5
START
Error message output
Stop ARARM/UNLOAD CASS.
Alarm disabled
Cassette removed
Subsequently, degeneration is performed
for the lamp and the IP is left unerased ***.
END
CR-IR341 Service Manual
7
8
9
10
Process continued
FR1H0445.EPS
Process continued
FR1H0446.EPS
MD - 158
MD - 158