12.2 Removal and Reinstallation of Erasure Conveyor
Removal
(1) Remove the erasure lamp assembly.
"12.1 Removal and Reinstallation of Erasure Lamp Assembly"
(2) Disconnect the connectors (CNC1, CNF1,2, CNSOLF1).
(3) Slid the tensioner until the timing belt can be removed.
Loosen the one retaining screw and slide the tensioner as appropriate, and then
tighten the screw.
"3.4 Tensioner Not Requiring Tension Adjustment" in Machine Description
(4) Remove the timing belt.
(5) Remove the two screws that retain the erasure conveyor, and take out the erasure
conveyor horizontally.
(2)
CNF1
009-051-04
05.31.99 FM2460
13
14
15
16
(2)
CNSOLF1
Erasure conveyor
(5)
(5)
Erasure conveyor
Spacer x2
BR M4x8 x2
CR-IR341 Service Manual
(2)
CNF2
Before-side-positioning
Conveyor
MC - 220
(4)
Timing belt
(3)
Tensioner
BR M4x8
(2)
CNC1
FR1H1703.EPS