FujiFilm CR-IR341 Service Manual page 215

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
5.11
Cassette Hold Release Operation
(1) Overall/General Sequence
Fig. 5.11.1 Initialization Cassette Hold Release Operation
Preprocessing: Fig. 5.1.1
Check for
next shelf
Check for
initialization
cassette
hold release
009-051-04
05.31.99 FM2460
3
4
5
6
START
Cassette hold release
DRIVE (SOLA1/ON)
DELAY TA51
Is SA3 opened?
Y
N
Is SA3 opened within TA11?
N
Y
Y
13A3
Processing
for
cassette
hold
release
failure
Y
03A3
Released for all shelves?
N
Y
END
If the shelf intended for ejection is degenerated due to initialization,
another empty cassette is put into another shelf so that the process
resumes.
CR-IR341 Service Manual
7
8
9
10
Table 5.11.1 Operation Parameters
Symbol
TA11
TA12
TA51
NA11
Table 5.11.1a List of I/O Names Used in the Flow and Drawing
I/O name
in flow
SA3
SOLA1
Is the number of retries made greater
than the retry count (NA11)?
N
23A2
Level 1 error
Cassette shelf error
Message display
"CASS. SET
INCORRECT CS#."
Degeneration for
relevant shelf
A
Degeneration for all shelves?
N
Level 0 error
Degeneration for all shelves
MD - 152
Description
Hold release timeout
Hold release retry
Wait for initialization hold
Hold release retry
I/O name
Topmost
2nd shelf
3rd shelf
shelf
position
position
position
SA3
SA7
SA11
SOLA1
SOLA2
SOLA3
Level 2 error
Cassette hold release
failure retry
Cassette hold
DRIVE (SOLA1/OFF)
DELAY TA12
Value
0.5 sec
2 sec
0.5 sec
5 times
TR1H0090.EPS
4th shelf
position
SA15
SOLA4
TR1H0091.EPS
Retry at
the time
of cassette
hold release
failure
FR1H0439.EPS
MD - 152