FujiFilm CR-IR341 Service Manual page 593

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
Checking and replacing shock absorbers of before-side-positioning
changeover guide and before-side-positioning retention guide
(1) Remove the before-side-positioning conveyor.
"6.1 Removing and Reinstalling Before-Side-Positioning Conveyor"
(2) Remove the guide plate.
(3) Check if the shock absorbers on the before-side-positioning changeover guide and
before-side-positioning retention guide.
If it is peeled or frayed, proceed to step (4) to replace the whole guide plate to which
the shock absorber is attached. If it is not peeled or frayed, proceed to step (14).
Before-side-positioning
(4) Remove the spring.
(5) Remove the arm.
(6) Remove the bearing.
(7) Remove the shock absorber-attached before-side-positioning changeover guide.
009-051-04
05.31.99 FM2460
3
4
5
6
conveyor
Before-side-positioning conveyor
CR-IR341 Service Manual
7
8
9
10
(3)
Shock absorber-attached
before-side-positioning
changeover guide
(3)
Shock absorber-attached
before-side-positioning
changeover guide
(2)
Guide plate
BR M3x6, x4
(5)
Arm
Sems M3x10
(4)
Spring
(7)
Shock absorber-attached before-side-
positioning changeover guide
MC - 102
FR1H1550.EPS
(6)
Bearing
BR M4x12, x2
FR1H1551.EPS
MC - 102