(8) Install a new before-side-positioning changeover guide in reverse order of removal
steps (4) through (7).
When installing the arm, tighten the screw with the arm pushed against the flat ma-
chined surface.
(9) Adjust the mounting position of the convergence path changeover guide drive solenoid
(SOLC1) so that the distance between the side face of the before-side-positioning
changeover guide and the back of the guide plate is 1 mm or less.
Before-side-positioning
changeover guide
009-051-04
05.31.99 FM2460
3
4
5
6
(8)
Arm
Sems M3x10
Flat machined surface
Guide plate
(363N1758B)
CR-IR341 Service Manual
7
8
9
10
FR1H1552.EPS
1 mm or less
FR1H1547.EPS
MC - 103
MC - 103