FujiFilm CR-IR341 Service Manual page 610

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
8.6
Replacing Grip Release Motor (MD2)
Replacement
(1) Remove the side-positioning conveyor.
"8.1 Removing and Reinstalling Side-Positioning Conveyor"
(2) Cut the cable tie.
(3) Disconnect CNMD2.
(4) Remove MD2 together with the bracket.
(5) Remove MD2 from the bracket.
(6) Remove the timing belt wheel and stay from MD2.
(5)
BR M4x8
Side-positioning
conveyor
(7) Install to a new MD2 the timing belt wheel that was removed at step (6).
(8) For reinstallation, reverse the removal steps.
Put the timing belt onto the timing belt wheel of MD2 and then secure MD2 in place.
009-051-04
05.31.99 FM2460
3
4
5
6
(6)
Stay
Pan-head
M4x10, x2
(6)
Timing belt wheel
(4)
Pan-head
M4x10, x2
CR-IR341 Service Manual
7
8
9
10
MD2
(3)
CNMD2
MC - 119
FR1H1781.EPS
MC - 119