FujiFilm CR-IR341 Service Manual page 897

Fuji computed radiography
Table of Contents

Advertisement

C
11
12
ONTENTS
11.11 5. UNIT: 7. DRIVEN GRIP: Operation Tests on Subscanning
Driven Grip Mechanism
1. GRIP: To Effect Driven Shaft Grip Operation
(1) Execute "1. GRIP".
The following message appears.
INPUT THE NIPPING SPEED.
1. HIGH
2. LOW
1 - 2 :
Meaning: Enter the grip speed.
(2) Type in either "1" or "2".
The following message appears.
MOVED TOWARD THE NIPPING POSITION IS IN PROGRESS.
DRIVE TIME : XXX
RESULT - OK
Meaning: Grip operation of the driven shaft is now being performed.
DRIVE TIME is displayed in units of 10 msec.
If it ends in error, "ERROR (XXXX)" appears. "XXXX" denotes an error code.
2. RELEASE: To Release Driven Shaft Grip
(1) Execute "2. RELEASE".
The following message appears.
INPUT THE NON-NIPPING SPEED.
1. ST
2. HR
1 - 2 :
Meaning: Enter the release speed.
(2) Type in either "1" or "2".
The following message appears.
MOVED TOWARD THE NON-NIPPING POSITION IS IN PROGRESS.
DRIVE TIME : XXX
RESULT - OK
Meaning: Grip release operation of the driven shaft is now being performed.
DRIVE TIME is displayed in units of 10 msec.
If it ends in error, "ERROR (XXXX)" appears. "XXXX" denotes an error code.
009-051-05
11.20.99 FM2638
13
14
15
16
CR-IR341 Service Manual
17
MU - 106
MU - 106