C
1
2
ONTENTS
IP Load Leak
After the suction cup is moved toward the cassette cover by 88 mm (262 pulses), the IP
sucked is detached from the suction cup as SVB1 is activated for air leak.
Because the movement speed of the suction cup is the same as the rotation speed of the
feed roller, the trailing edge of the IP is inserted into the cassette.
The IP sucked is moved to
the cassette cover and detached.
009-051-04
05.31.99 FM2460
3
4
5
6
Cassette inner frame
when the cover is closed
Unit: mm
23.83
5.5
IP accommodation position
(trailing edge when IP is placed
at the bottom of the cassette)
CR-IR341 Service Manual
7
8
9
10
FR1H0210.EPS
MD - 23
Atmospheric
pressure
SVB 1
PB 1
OFF
OPEN
During IP leak
MD - 23
ON
SB5
IP
FR1H0211.EPS