11
12
ONTENTS
11.2 2. MOTOR: Operation Tests on Motor
It is used during checking of the pulse motors.
1. NUMBER: To List Up No. of Pulse Motor
Execute "1. NUMBER".
A list of pulse motor numbers is displayed as follows.
1:MB1 2:MB2 3:MB3 4:MC1 5:MD1
6:MD2 7:MD3 8:ME1 9:ME2 10:MF1
2. PARAMETER: To Set Drive Parameter
(1) Select "2. PARAMETER".
The following message appears.
1:MB1 2:MB2 3:MB3 4:MC1 5:MD1
6:MD2 7:MD3 8:ME1 9:ME2 10:MF1
INPUT THE NUMBER OF MOTOR.
1 - 10 :
Meaning: Enter the number of motor.
(2) Type in the number of motor from 1 through 10.
The following message appears.
INPUT THE ROTATION DERECTION.
1.CW 2.CCW (FIXED=X)
Meaning: Enter the rotation direction.
"X" denotes the rotation direction (either "1" or "2") that is currently set.
(3) Type in the rotation direction (either "1" or "2").
The following message appears.
INPUT THE TOTALNUMBER OF PULSES.
0 - 8388607(FIXED=XXXXXXX) :
Meaning: Enter the total number of pulses.
"XXXXXXX" denotes the total number of pulses that is currently set.
(4) Type in the total number of pulses.
The following message appears.
INPUT THE HI-SPEED.
16 - 7200 (FIXED=XXXX) :
Meaning: Enter the high-speed value.
"XXXX" denotes the high-speed value that is currently set.