5.10 Replacing IP Transport Motor (MB2)
(1) Rotate the timing belt until the IP removal unit is moved up (away about 150 mm from
the tray of the cassette set unit) to facilitate MB2 removal.
(2) Remove the cable clamp from the connector (CNMB2).
(3) Disconnect the connector (CNMB2).
(4) Take out MB2.
(5) Remove the timing belt pulley and wheel from the MB2 taken out, and attach them to a
new MB2.
E-ring
(5)
Timing belt wheel
(6) Secure MB2 in place with two screws.
At this time, put the timing belt onto the timing belt wheel, and fit the protruding portion
of the motor into the hole of the bracket.
(7) Connect the connector (CNMB2).
(8) Put the cable clamp back where it was.
009-051-04
05.31.99 FM2460
3
4
5
6
Bracket
(4)
MB2
Pan-head
M4x10, x2
CR-IR341 Service Manual
7
8
9
10
(1)
Timing belt
MB3
(2), (3)
CNMB2
(5)
Wheel
Hex socket head setscrew M4x6