10.14 Replacing Grip Arms
(1) Remove the rubber roller located above the subscanning unit.
"10.13 Replacing Rubber Roller Located Above Subscanning Unit"
(2) Remove the light-collecting mirror.
"10.4 Servicing Light-Collecting Mirror"
(3) Remove MZ1 (FFM).
"10.8 Checking and Replacing Subscanning Motor (MZ1)"
(4) Remove the flywheel.
"10.12 Checking and Replacing Kapton
of flywheel"
(5) Remove the right-hand side transparent cover.
(6) Remove the two screws that retain each of the two grip arms.
"10.15 Replacing Guide Plate of Subscanning Unit"
(7) Take out the grip arm from the subscanning unit.
(8) Remove the bearing and shaft from the grip arm.
Hex socket head setscrew
(9) Attach the bearing and shaft removed at step (8) to a new grip arm.
(10) Reinstall the grip arm to the subscanning unit.
(11) Reinstall the right-hand side transparent cover.
(12) Reinstall the flywheel.
"10.12 Checking and Replacing Kapton
of flywheel"
(13) Reinstall the MZ1 (FFM).
"10.8 Checking and Replacing Subscanning Motor (MZ1)"
(14) Reinstall the light-collecting mirror.
"10.4 Servicing Light-Collecting Mirror"
(15) Reinstall the rubber roller located above the subscanning unit.
"10.13 Replacing Rubber Roller Located Above Subscanning Unit"
009-051-04
05.31.99 FM2460
3
4
5
6
Grip arm
M4x8
CR-IR341 Service Manual
7
8
9
10
®
Belt/Tensioner/Flywheel,
Grip arm
Hex socket head setscrew
M4x8
Shaft
Bearing
®
Belt/Tensioner/Flywheel,
MC - 201
Replacement
FR1H1544.EPS
Replacement
MC - 201