Replacing Vertical Conveyor Guide - FujiFilm CR-IR341 Service Manual

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS

5.15 Replacing Vertical Conveyor Guide

Perform the procedures from the rear of the machine.
(1) Remove the idler gear from the vertical conveyor and then the timing belt.
(2) Remove the six grip springs located on both right and left of the vertical conveyor.
Do not remove the lowermost grip spring. Also, keep in mind that the right- and left-
hand grip springs differ in length.
(3) Remove the gear.
"3.1 E-ringless Housing Employed " in Machine Description
(4) Remove the housing and bearing.
Remove the right and left housings and bearings of the vertical conveyor that retain
the rubber roller in place. Do not remove the housing and bearing located at the
bottom.
R E
O N
O F
009-051-04
05.31.99 FM2460
3
4
5
6
Timing belt
A
A
A
(1)
Idler gear
W
T S
t e
S E
m o
l
c a
R e
L o
F
CR-IR341 Service Manual
7
8
9
10
(2)
Free length: 80 mm grip spring, x3
(2)
Free length: 93 mm grip spring, x3
MC - 84
(4)
Bearing
Vertical conveyor guide
(4)
Housing
(3) Gear
MC - 84
FR1H1075.EPS