FujiFilm CR-IR341 Service Manual page 274

Fuji computed radiography
Table of Contents

Advertisement

1
2
3
4
C
ONTENTS
Code
Name
0821
DSP20 memory read error
0822
DSP20 checksum error
0823
DSP20 self-diagnostics error
0824
DSP20 image processing error
0825
DSP21 device initialization error
0826
DSP21 µ load error
0827
DSP21 memory write error
0828
DSP21 memory read error
0829
DSP21 checksum error
082A
DSP21 self-diagnostics error
082B
DSP21 image processing error
082C
DSP22 device initialization error
082D
DSP22 µ load error
082E
DSP22 memory write error
082F
DSP22 memory read error
009-051-04
05.31.99 FM2460
5
6
Significance/Occurrence Condition
Read from the DSP device cannot be done.
DSP µ checksum error
DSP µ self-diagnostics error
DSP µ image parameter error
The DSP device cannot be opened.
The DSP µ is not yet registered in the HD.
Write to the DSP device cannot be done.
Read from the DSP device cannot be done.
DSP µ checksum error
DSP µ self-diagnostics error
DSP µ image parameter error
The DSP device cannot be opened.
The DSP µ is not yet registered in the HD.
Write to the DSP device cannot be done.
Read from the DSP device cannot be done.
CR-IR341 Service Manual
Detail: Denotes the detail information format.
Probable Cause/Remedy
Check the IMG08B board.
• Check the IMG08B board.
• Register the DSP µ in the HD again.
• Check the IMG08B board.
• Register the DSP µ in the HD again.
Check the image parameter.
Check the IMG08B board.
Register the DSP µ in the HD again.
Check the IMG08B board.
Check the IMG08B board.
• Check the IMG08B board.
• Register the DSP µ in the HD again.
• Check the IMG08B board.
• Register the DSP µ in the HD again.
Check the image parameter.
Check the IMG08B board.
Register the DSP µ in the HD again.
Check the IMG08B board.
Check the IMG08B board.
MT - 15
Detail
Refer to
I-7
I-7
I-7
I-7
I-7
I-7
I-7
I-7
I-7
I-7
I-7
I-7
I-7
I-7
I-7
TR1H1229 .EPS
MT - 15