03DE Elevation Movement Error
Cause
SB3 does not turn ON or OFF during elevation HP movement.
Supplementary explanation
During initialization, MD3 is driven to move the elevation unit, thereby turning ON
SA17.
As SA17 turns ON, the HP of the elevation unit is detected.
When SA17 turns ON, the elevation unit is subsequently moved for detection of SB3
(shelf 1).
The stop position of the elevation unit is determined as SB3 detects a transition from
ON to OFF to ON.
This error occurs when SB3 does not turn ON, or when it does not transition from ON
to OFF to ON, during the above operation.
Sensor related
Error-related I/O locations
Error-related system block diagram
AC IN
DC OUT2
JPS-1
DC OUT3
009-051-04
05.31.99 FM2460
3
4
5
6
ON: Close, OFF: Open
SA17
Elevation
HP sensor
SNS08A
CN
CN7
CN13
CN4
DRV03D
CN3
CN3
CN6
DRV08A
CN1
CR-IR341 Service Manual
SB3
Each shelf
position sensor
CNB1
CNB2
CNB4
MB3
CNMD3
FR1H0099.EPS
MT - 132
MB3
Up-down drive motor
FR1H0098.EPS