Setup Example of Conveyance Test Operation Mode
An example of how to set the operation mode for checking the mechanism of the IP convey-
ance function is presented below.
NOTE
When "2. AUTO MODE" is selected, be sure to set the operation mode of routine processing
(1. ROUTINE) before executing the routine processing. If the routine processing is executed
without this setting, operation error may result.
(1) Select "3. TEST MODE".
(2) Select "2. AUTO MODE".
The operation modes are displayed.
(3) Select "3. SECONDARY ERASURE".
A prompt asking you to enter the number of times appears.
(4) Enter 5 (the number of times it is executed).
(5) Select "0. QUIT" to return to the U-Utility screen.
(6) Touch the "RETURN" button.
The mode returns to the routine screen.
(7) Put the cassette into the machine.
IPs are conveyed the number of times that was set in step (4).