FujiFilm CR-IR341 Service Manual page 208

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
5.9
Subscanning Grip HP Positioning
(1) Overall/General Sequence
Fig. 5.9.1 Subscanning Grip HP Positioning
Preprocessing: Fig. 5.1.1
Misalignment
compensation
warm-up
operation
HP
positioning
Self-
diagnostics
Idle
positioning
Table 5.9.1 Operation Parameters
Description
Symbol
TZ51
Wait for initialization grip operation
009-051-04
05.31.99 FM2460
3
4
5
6
START
Driving shaft grip
release (fast)
Fig. 5.9.2
DELAY TZ51
Driving shaft grip (ST)
Fig. 5.9.3
DELAY TDZ51
Driving shaft grip
release (fast)
Fig. 5.9.2
DELAY TZ51
Driving shaft grip (HR)
Fig. 5.9.3
DELAY TZ51
Driving shaft grip
release (HR)
Fig. 5.9.2
DELAY TZ51
Driving shaft self-diagnostics
Fig. 5.9.7
DELAY TZ51
Driving shaft grip
release (fast)
Fig. 5.9.2
DELAY TZ51
A
Value
0.5 sec
TR1H0080.EPS
CR-IR341 Service Manual
7
8
9
10
A
Driven shaft grip
release (ST)
Fig. 5.9.4
DELAY TZ51
Driven shaft grip
(fast)
Fig. 5.9.5
DELAY TZ51
Driven shaft grip
release (ST)
Fig. 5.9.4
DELAY TZ51
Driven shaft grip
(HR)
Fig. 5.9.5
DELAY TZ51
Driven shaft
self-diagnostics
Fig. 5.9.6
END
MD - 145
Misalignment
compensation
warm-up
operation
HP
positioning
FR1H0430.EPS
MD - 145