12.14 Checking and Replacing Branch Path Changeover Guide
Check
CAUTION
Before removing the covers, be sure to turn OFF the high-voltage switch. If the machine is
powered ON with any of the covers removed, the photomultiplier will be damaged.
(1) Power ON the machine.
(2) Touch the U-Utility button.
The screen switches to the U-Utility mode.
(3) Touch the upper left and upper right corners of the operation panel in sequence.
M-Utility is activated.
(4) Select "6. MECHANICAL UTILITY", "2. ACTUATOR" and "2. DRIVE" in sequence.
A list of actuator Nos. is displayed.
(5) Select "7.SOLF1".
The branch path changeover guide is driven to the front of the machine.
If the branch path changeover guide is not driven, it is likely that the cable connection
or sensor is faulty. So, check the cable connection, and then select "2. DRIVE" again.
If it is not driven after that, replace SOLF1.
(6) Repeatedly select "0. QUIT".
M-Utility is exited.
009-051-04
009-051-04
05.31.99 FM2460
05.31.99 FM2460 (1)
13
14
15
16
Branch path
changeover guide
Changeover guide
drive solenoid
CR-IR341 Service Manual
Erasure conveyor
FR1H1258.EPS
MC - 240