FujiFilm CR-IR341 Service Manual page 556

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
(9) Rotate the MB3 timing belt wheel clockwise, as seen from the rear, until the actuator
blocks the SB3 light path.
A message indicating that the sensor is CLOSE appears on the operation panel.
(10) Repeat steps (8) and (9) several times.
If anything abnormal is found, such as no message displayed, then replace the sensor.
(11) Repeatedly select "0. QUIT".
M-Utility is exited.
Replacement (replacement should be performed without removing the bracket
where possible)
If the SB3-attached bracket is removed during replacement, be sure to adjust SB3.
(1) While holding it up, remove the each shelf position sensor (SB3).
"3.5 Sensors Not Requiring Adjustment of Its Mounting Position" in Machine
Description
(2) Disconnect the connector (SB3).
(3) Connect the connector (SB3).
(4) Install a new sensor.
(5) Check SB3.
" Checks"
Adjustment
(1) Loosen the screw on the SB3 mounting bracket.
(2) Tighten the screw on the SB3 mounting bracket in place so that the distance between
the lower portion of the side plate of the IP removal unit and the bottom of SB3 is 99.5
to 100.0 mm.
(3) Check SB3.
" Checks"
009-051-05
11.20.99 FM2638
3
4
5
6
IP removal unit
SB3
99.5 — 100.0mm
CR-IR341 Service Manual
7
8
9
10
(2)
Bracket
Connector
Latch
(1)
SB3
Bottom
of SB3
FR1H1768.EPS
MC - 65
FR1H1065.EPS
MC - 65