FujiFilm CR-IR341 Service Manual page 294

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
3
4
ONTENTS
Code
Name
Suction cup HP detection preparation
23E6
retry
23E7
Suction cup HP detection retry
23E8
Suction cup HP return error
23E9
IP leak error
Cassette detection logic failure 1
23F1
(immediately after cassette setting)
Cassette detection logic failure 2
23F2
(after up-down movement)
23F4
Feed conveyance motor error
23F5
Side-positioning conveyance motor error
23F6
Before-erasure conveyance motor error
23F8
After-erasure conveyance motor error
23F9
NFB trip detection
009-051-05
009-051-06
04.20.2001 FM3006
11.20.99 FM2638
5
6
Significance/Occurrence Condition
A leak error occurs during IP loading.
MB3 is stopped during IP length measurement conveyance,
but the pulse motor stop interrupt is not detected after a
predetermined period of time has elapsed.
MB3 is stopped during side-positioning conveyance, but the
pulse motor stop interrupt is not detected after a predet
ermined period of time has elapsed.
MF1 is stopped during before-erasure conveyance,
but the pulse motor stop interrupt is not detected after a
predetermined period of time has elapsed.
MB2 is stopped during IP length measurement conveyance,
but the pulse motor stop interrupt is not detected after a
predetermined period of time has elapsed.
1 is written to the port on the SNS board, but it is 0
during reading.
CR-IR341 Service Manual
Detail: Denotes the detail information format.
Probable Cause/Remedy
Detail
MT - 33
Refer to
MD - 63
D-5
MD - 64
D-5
MD - 62, 65, 69,
D-5
93, 101, 111, 114,
123
D-5
D-5
D-5
D-5
D-5
D-5
D-2
TR1H1247.EPS
MT - 33