FujiFilm CR-IR341 Service Manual page 120

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
2.
Routine (Single-IP Conveyance)
2.1 Summary of Operation
(1) Machine Operation during Single-IP Conveyance (Overall/General Flow)
The machine operation during single-IP conveyance is as follows.
The general relationship between unit components and processes along the conveyance
paths is illustrated on the next page.
Fig. 2.1.1 Machine Operation during Single-IP Conveyance
Cassette set unit
In this sequence, the hold on the cassette inserted is detected
and confirmed.
Up-down IP removal unit
In this sequence, the elevation unit is moved to the shelf where
the cassette that loads the IP has been set.
Up-down IP removal unit
In this sequence, the IP is removed from the cassette and
conveyed to the before-side-positioning conveyor.
Before-side-positioning conveyor
In this sequence, the IP from the up-down IP removal unit is
conveyed to the side-positioning conveyor. During the conveyance,
IP length measurement is performed.
After the IP length measurement, the erasure lamp turn-on process
is performed.
Side-positioning conveyor
In this sequence, the barcode on the back of the IP is read during
the conveyance; subsequently, positioning of the start point for IP
reading (i.e., side-positioning) is performed.
Subscanning unit
In this sequence, after the IP data is read, the IP is conveyed to the
conveyance-inverting stocker.
Up-down IP removal unit
In this sequence, the elevation unit is moved to the shelf where the
cassette that loads the IP has been set.
Erasure conveyor
The IP in the inverting stocker is passed under the erasure lamp
(data is erased) by switching the conveyance path of the erasure
conveyor, and conveyed to the before-side-positioning conveyor.
The erasure lamp turn-off process is also performed.
Before-side-positioning conveyor
The IP sent from the erasure conveyor is conveyed to the up-down
IP removal unit by switching the conveyance path, and loaded into
the cassette.
Cassette set unit
In this sequence, the hold on the cassette into which the IP has
been loaded is released, so that the cassette
009-051-04
05.31.99 FM2460
3
4
5
6
After-reading conveyor
Up-down IP removal unit
CR-IR341 Service Manual
7
8
9
10
START
See 2.2
See 2.3
See 2.4
See 2.5
See 2.6
See 2.7
See 2.9
See 2.10
See 2.11
END
MD - 57
Cassette IN
Movement of elevation unit (feed)
IP feed
Feed conveyance (IP length measurement)
Barcode reading and side-positioning
Reading
Movement of elevation unit (load)
See 2.8
Erasure
IP load
Cassette ejection
FR1H0301.EPS
MD - 57