FujiFilm CR-IR341 Service Manual page 395

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
03D9 Erasure Conveyance Remaining IP Ejection Error
Because SF2 is ON during initialization, the IP is conveyed to the erasure conveyor,
but SF3 does not turn ON.
Or, because SF3 is ON, the IP is conveyed to the cassette set unit, but SF3 does not
turn OFF.
Supplementary explanation
During initialization, each motor is activated, and each sensor is used to search for
any remaining IP left in the machine.
The remaining IP is detected by SF4, SF2, and SF3 of the erasure conveyor, so the
IP ejection process is performed.
When SF2 detects the IP, MF1 is driven to convey the IP toward the erasure lamp
unit, thereby causing it to pass over SF3. (SF3: OFF to ON)
When SF3 detects the IP, MF1 is driven to convey the IP toward the cassette, thereby
causing it to pass over SF3. (SF3: ON to OFF)
This error occurs when the IP fails to pass over SF3 (SF3 status does not transition
from OFF to ON), or when it does not completely pass over SF3 (SF3 status does not
transition from ON to OFF), during the above operation.
Sensor related
Error-related I/O locations
IP transport motor
Changeover IP sensor
Error-related system block diagram
AC IN
DC OUT2
JPS-1
DC OUT3
009-051-04
05.31.99 FM2460
3
4
5
6
ON: Close, OFF: Open
MF1
SF2
SNS08A
CN
CN8
CN13
CN4
DRV03D
CN3
CN3
CN10
DRV08A
CN1
CR-IR341 Service Manual
SOLA3
SOLA4
SF3
Load standby IP sensor
CNF1
CNF2
MF1
CNMF1
FR1H0089.EPS
MT - 126
FR1H0088.EPS
MT - 126