12.7 Replacing IP Transport Motor (MF1)
(1) Remove the erasure conveyor.
"12.2 Removal and Reinstallation of Erasure Conveyor"
(2) Cut the cable tie.
(3) Disconnect the connector (CNMF1).
(4) Remove the connector from the side plate.
Because the connector portion of the motor is secure by a latch onto the side plate of
the erasure conveyor, push down the latch to remove the connector.
(5) Remove MF1.
(6) Using an Allen wrench, remove the set screw, and take out the wheel from MF1.
(7) Remove the E-ring and take out the timing belt wheel.
(6)
Wheel
Hex socket head setscrew
M4x4
(8) Reverse the removal steps to reinstall a new MF1.
009-051-04
05.31.99 FM2460
13
14
15
16
Bearing
(9)
Timing belt
(4)
Connector
(8)
Bracket
BR M4x8 x3
(7)
E-ring
(7)
Timing belt wheel
(5)
MF1
Pan M4x10 x4
CR-IR341 Service Manual
Erasure conveyor
(3)
CNMF1
MC - 230