FujiFilm CR-IR341 Service Manual page 823

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
Item
IDT
Item
number
type
HQ_
48
FUNC
DRC_
49
FUNC
ID_
50
INFO_
TYPE
ID_
51
CODE_
SET
52
SSS
DST_
53
PADDING
54
ENV
ID_
55
RET
56
UID
009-051-07
08.30.2001 FM3115
3
4
5
6
Initial value
Setup description
HQ function setting
1
DRC function setting
1
ID# input type for CSL type
0
Acceptable character set during
0
ID online for CSL type
0
Screen saver selection
2
Distribution code padding setting
Machine type display switching
CR-IR341
setting
Patient information acquisition
0
setting during ID online
0
Study/Series information
numbering setting
CR-IR341 Service Manual
7
8
9
10
Parameters
0: disabled; 1: enabled
0: disabled; 1: enabled
0: Stand.
1: Long
0: CP850
1: ISO8859-1
0: Blank screen
(Screen saver is disabled)
1: Pictures
(Multiple pictures are displayed
at regular intervals.)
0: Pad with space to the head
(EX."
1234")
1: Pad with zero to the head
(EX."00001234")
2: Pad with space to the tail
(EX."1234
3: Pad with zero to the tail
(EX."12340000")
0: The patient ID is automatically
cleared when the cassette is set
or when the menu select key is
pressed, and the patient
information is acquired from the
online destination.
1: The patient ID is cleared when
it is necessary to change the
patient ID, and the patient
information is acquired from the
online destination.
<CSL type>
0: Study/Series information is not
numbered on the reader side.
1: Study/Series information is
numbered on the reader side.
<IDT type>
0: Study/Series information is not
numbered on the IDT side.
1: Study/Series information is
numbered on the IDT side.
MU - 40
Remarks
No change
required
except
for USA
No change
required
except
for USA
")
TR1B1107.EPS
MU - 40

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir341pFcr 5000Fcr 5000plus

Table of Contents