FujiFilm CR-IR341 Service Manual page 597

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
7.
Pulling Out and Pushing In Scanner Unit
Pulling Out
CAUTIONS
• Before pulling out the scanner unit, be sure to remove the two bracket retaining screws lo-
cated on the upper portion of the SCN08A board and disconnect the connectors. If pulled out
without removing them, the SCN08A board may be damaged.
• When pulling out the scanner unit, be sure to lower the adjustable feet down to the floor. If not,
the bolts of the adjustable feet would interfere with the cable cover.
• When servicing any printed circuit board, be sure to wear an anti-static wristband to ensure
proper grounding. If not, static electricity built on the human body may cause damage to
electric parts mounted on the board.
(1) Remove the following covers in order named.
Lower front cover, lower right-hand side cover
"3. Removing and Reinstalling Covers"
(2) Disconnect the connectors (CN7-9, CNZ6-8, 10 on the SCN08A board, and CN1, 2, 4,
5 on the subscanning unit).
Subscanning unit
(2)
CNZ4
CNZ5
009-051-04
05.31.99 FM2460
3
4
5
6
CNZ2
CNZ1
Scanner unit
CR-IR341 Service Manual
7
8
9
10
SCN08A board
(2)
CN7
CN9
MC - 106
Remove screw
(BR M4x8, x2)
CN8
CNZ6
CNZ7
CNZ8
CNZ10
FR1H1555.EPS
MC - 106