4.12 Replacing Rubber Rollers of Cassette Set Unit
(1) Pull out the tray of the cassette set unit.
"4.3 Removing and Reinstalling Shutter Assembly"
(2) Remove the two grip springs of the rubber roller to be removed.
(3) Remove the two bearings and two housings.
(4) Replace the rubber roller.
(5) For reinstallation, reverse the removal steps.
009-051-04
05.31.99 FM2460
3
4
5
6
(2)
Grip spring
(3)
Housing
(3)
Bearing
(4)
Rubber roller
CR-IR341 Service Manual
7
8
9
10
(1)
Cassette set unit
MC - 55
(3)
Bearing
(3)
Housing
(2)
Grip spring
FR1H1755.EPS
MC - 55