FujiFilm CR-IR341 Service Manual page 619

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
Replacement
(1) Remove the side-positioning conveyor.
"8.1 Removing and Reinstalling Side-Positioning Conveyor"
(2) Remove MD1 together with the bracket.
"8.5 Replacing Side-Positioning Motor (MD1)"
(3) Remove the retainer that retains the timing belt.
(4) Remove the timing belt.
T
O
P
(5) Install a new timing belt.
(6) Reinstall the retainer to the timing belt.
(7) Reinstall MD1.
"8.5 Replacing Side-Positioning Motor (MD1)"
(8) Adjust the timing belt.
" Adjustment"
009-051-04
05.31.99 FM2460
3
4
5
6
(2)
MD1
CR-IR341 Service Manual
7
8
9
10
(3)
Retainer
BR M4x8, x2
(4)
Timing belt
MC - 128
FR1H1791.EPS
MC - 128

Advertisement