6.8
Replacing Before-Side-Positioning Conveyor Rubber
Rollers
(1) Remove the before-side-positioning conveyor.
"6.1 Removing and Reinstalling Before-Side-Positioning Conveyor"
(2) Remove the grip spring for the rubber roller to be replaced.
(3) Remove the drive gear.
"3.1E-ringless Housing Employed" in Machine Description
(4) Remove the guide plate.
"6.9 Checking and Replacing Shock Absorber-Attached Guide of Before-Side-
Positioning Conveyor"
(5) Remove the bearing and take out the rubber roller.
Side view
(6) For reinstallation, reverse the removal steps.
NOTES
When installing the rubber rollers, observe the following precautions.
• Ensure that the flat machined surfaces of a pair of rubber rollers face each other.
• Be sure that the Bearing rotation prevention bracket is positioned toward the 20 mm
diameter roller.
Bearing
• When installing the grip spring, exercise care not to erroneously orient it because its
left-hand portion is longer than the right-hand portion.
009-051-04
05.31.99 FM2460
3
4
5
6
: Rubber roller
Bearing
CR-IR341 Service Manual
7
8
9
10
65 mm diameter roller
Flat machined surface
Bearing rotation
prevention bracket
Flat machined surface
20 mm
diameter roller
MC - 99
Before-side-positioning
conveyor
FR1H1116.EPS
FR1H4013.EPS
MC - 99