FujiFilm CR-IR341 Service Manual page 821

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
I/F Setup (Item Numbers 29-82)
Item
IDT
Item
number
type
29
IFT
30
PRO
31
SF1
32
SF2
33
GRP
34
PTR
35
UNQ
36
PCD
37
PCL
38
CIM
39
EBK
40
EBA
009-051-07
08.30.2001 FM3115
3
4
5
6
Initial value
Setup description
0
I/F type
Function image protect
(image protect on HIC)
0
Use to set the type of screen
that enables the screen protection
function of HI-C.
Sensitivity median shift 1
50
(relatively high sensitivity)
Sensitivity median shift 2
100
(high sensitivity)
Group information setup method
0
for sorting
Position of film mark or menu
character string outputted with
0
GRP (valid when GRP = 2, 3, 6)
Unique information setup method
0
for sorting
1
Clock during DMS-I/F output
0
Clock during LP-I/F output
DMS output code selection when
0
"inch" and "metric" physical sizes
differ
Selection of a menu to transmit
EDR backup information to the IP
0
address that is set with EBA
Transfer destination IP address for
the transmission of EDR backup
. . .
data through a network
The transmission port is fixed
at 90E.
CR-IR341 Service Manual
7
8
9
10
RESERVE
0: NONE
1: TS (Temporary subtraction)
2: ES (Energy subtraction)
3: MS (Multilayer tomography)
4: MX (Multi-integrated processing)
5: WS (Whole spine)
6: PS (IVP)
7: TM (Tomography)
8: TX (Temporary subtraction,
overall processing)
9: WX (Whole spine, overall
processing)
10: EX (Energy subtraction, overall
processing)
11: AS (Inter-image operation)
99: ALL
0 — 299
0 — 299
0: space (10)
1: space (2) + department name (8)
2: department name (4) + menu (6)
3: department name (4) + film mark (6)
4: unit code (1) + space (9)
5: space (4) + MPM (4) + space (2)
6: space (4) + film mark (6)
7: department name (4) + MPM (4) +
space (2)
0 — F (hexadecimal notation)
0: Space (10)
1: patient ID number (10)
0: 1 sec
1: 2 sec
0: 1 sec (< 60m)
1: 2 sec (> 60m)
0: Inch
1: Metric
0: OFF (not transmitted)
1: MENU (menu selected with
EM is selected)
2: Full (all menus)
Transfer destination IP address
The IP address should be described
in fifteen digits stuffed to the left.
If the number of digits
is less than 15, space should be
added to the end to meet the 15-digit
requirement.
MU - 38
Parameters
Remarks
No change
required.
No change
required.
It should be
set
according to
the
processing
at the DMS.
MU - 38
No change
required.
Indent
Indent
Indent
LENGTH
=15
TR1B1008.EPS