FujiFilm CR-IR341 Service Manual page 232

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
Symbol
TF12
After-reading conveyance IP trailing-edge detection
timeout
TF13
Wait for switchback
TF14
Before-erasure conveyance IP leading-edge
detection timeout
TF15
Erasure conveyance IP trailing-edge detection
timeout
TF17
After-erasure conveyance IP trailing-edge detection
timeout 1
TF18
After-erasure conveyance IP trailing-edge detection
timeout 2
TF51
Wait for initialization erasure lamp turn-on
TF52
Wait for initialization erasure lamp turn-off
TF71
After-reading conveyance remaining IP ejection
timeout
TF72
Erasure conveyance remaining IP ejection timeout
TF73
Erasure conveyance remaining IP search timeout
TF74
Load standby positioning timeout 1
TF76
Reading conveyance remaining IP search timeout
TF77
Load standby positioning timeout 2
TQ21
SPR after-erasure conveyance timeout
TQ22
SPR load timeout
TR11
SPR before-erasure conveyance timeout
TR12
SPR erasure conveyance IP trailing-edge detection
timeout
009-051-04
009-051-04
05.31.99 FM2460 (1)
05.31.99 FM2460
3
4
5
6
Description
CR-IR341 Service Manual
7
8
9
10
Ref. value
(sec.)
14x17 inch
1.38
14x14 inch
1.22
10x12 inch
1.11
8x10 inch
0.89
0.84
1824 ST
1.10
2430 ST
1824 HR
0.84
2430 HR
0.97
0.2
0.4
14.2
1.2
1.3
5
75
5.6
4.8
3.9
1.2
4.8
5.6
2.6
3.7
0.42
0.59
0.69
0.91
0.96
0.7
0.96
0.83
MD - 169
Remarks
TR1H1156.EPS
MD - 169

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir341pFcr 5000Fcr 5000plus

Table of Contents