FujiFilm CR-IR341 Service Manual page 428

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
13E2/13E5
Preliminary steps:
1. Power OFF the machine and remove the covers.
2. Turn OFF the high-voltage switch (S1) of the SCN08A board.
3. Check for any foreign matter (lead character marking chips, etc.) inside the machine, and if any, remove it all.
Power OFF
Is the IP normal?
Make sure that its sucked surface
is not soiled, curled, or bent.
Y
Power OFF
Are the suction cups of the IP
removal unit normal?
Check if the suction cups are
installed properly.
Check the suction cups for any
soil or deformation.
Y
Power OFF
Is the air hose of the IP removal
unit normal?
Check if the air hose/hose joint
are installed properly.
Y
Power ON
Is SB5 normal?
With "6. MECH" of M-Utility,
execute sensor monitoring to
check the SB5 status.
Y
Restart the machine
and check its status.
009-051-04
05.31.99 FM2460
3
4
5
6
N
Clean or replace the IP.
N
Clean/adjust/replace the following parts.
Suction cup rubber
Suction cup retaining screw
N
Adjust/replace the following parts.
Air hose
Hose joint
Power OFF
N
Is SB5 installed normally?
Y
Power OFF
Is the connector connection
from SB5 to SNS08A board
normal?
Y
Replace the following parts
in order named.
SB5
SNS08A board
CR-IR341 Service Manual
N
Install the sensor properly.
N
Check the error-related system block
diagram and connect the connectors
as appropriate.
MT - 159
FR1H0130.EPS
MT - 159