FujiFilm CR-IR341 Service Manual page 872

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
2. FREQ ADJUST: To Adjust Read Width
The read width is adjusted. The setup value is not written to the hard disk.
(1) Select "2. FREQ ADJUST".
The following message appears.
INPUT MAIN SCAN LENGH ADJUSTMENT VALUE.
-5.00 - +5.00 :
(2) Enter the read width adjustment value over a range from -5.00 to 5.00 (in %).
Enter an incremental or decremental value in percentage relative to the current read
width.
<Example>
To increment the read width on the output screen from 172 mm to 175 mm:
(175 - 172)/172 x 100 ≈ 1.74
Thus, enter "1.74".
The result is displayed.
RESULT - XXXXXXXXXX
XXXXXXXXXX represents the result (OK: normal, ERROR: error)
3. PIXEL ADJUST: To Adjust Read Start Position
The read start position is adjusted. The setup value is not written to the hard disk.
(1) Select "3. PIXEL ADJUST".
The following message appears.
INPUT INVALID PIXELS.
-999 - 999 :
(2) Enter the number of pixels shifted for the read start position over a range from -999 to
999 (in pixels).
Ten pixels correspond to 1 mm in actual measurement on the IP.
To shift rightward on the output image, enter a positive value, and to shift leftward,
enter a negative value.
<Example>
Shifting the two-in-one image of 14"x17" leftward by 1 mm means that it is shifted
leftward by 2 mm on the IP because the reduction ratio is 1/2.
Thus, enter "-20".
The result is displayed.
RESULT - XXXXXXXXXX
XXXXXXXXXX represents the result (OK: normal, ERROR: error).