4.7
Checking and Replacing Mammo/ST Sensors (SA4, 8, 12,
16)
Checks
(1) Place the machine in a sensor monitoring state so that SA4, 8, 12, and 16 can be
monitored.
"2.2 Checking Various Sensors"
(2) Set a non-mammo-type cassette into the shelf in place.
A message indicating that the sensor is OPEN appears on the operation panel.
Cassette
(3) Pull out the cassette.
A message indicating that the sensor is CLOSE appears on the operation panel.
(4) Set a mammo-type cassette in place.
When the cassette is inserted, a message indicating that the sensor is OPEN
appears. Immediately after that, a message indicating that the sensor is CLOSE
appears.
(5) Pull out the cassette.
When the cassette is removed, a message indicating that the sensor is OPEN
appears. Immediately after that, a message indicating that the sensor is CLOSE
appears.
(6) Repeat steps (4) and (5) several times to ensure that the messages appear as appro-
priate.
If anything abnormal is found, such as no message displayed, then replace SA4, 8, 12,
16, and then check it again.
(7) Enter "0 (zero)".
The monitoring is exited to return to "3. SENSOR" menu.
009-051-04
05.31.99 FM2460
3
4
5
6
CR-IR341 Service Manual
7
8
9
10
Cassette set unit
MC - 40
SA4
SA8
SA12
SA16
FR1H4047.EPS