Adjustment
(1) Loosen the two respective retaining screws on the four shock-absorbing rubbers.
The screw that retains the stay of the subscanning unit and the shock-absorbing
rubber should not be loosened.
(2) Adjust the position of the subscanning unit so that its distances from the tray are as
indicated by adjustment values (A, B, and C) shown below.
(3) Tighten the screws that were loosened at step (1).
(4) Check again the distances from the subscanning unit to the tray.
If the adjustment values (A, B, and C) are out of range, repeat steps (1) through (3) all
over again.
009-051-04
05.31.99 FM2460
3
4
5
6
Shock-absorbing rubber x4
Loosen two screws
CR-IR341 Service Manual
7
8
9
10
Top view
FRONT
C
A = 74 – 1, B = 74 – 1, C = 100 – 1
MC - 206
Tray
A
Subscanning unit
B
Unit: mm
FR1H1615.EPS
MC - 206