ONTENTS
How to Check Fuse
For no-fuse breaker (NFB)
The no-fuse breakers (NFB1-NFB2) are to be shorted (N) at factory prior to shipment and
must be so at installation. However, since they may unexpectedly turn OFF due to vibration
during transit, press all the no-fuse breaker buttons two or three times to ensure that they
are depressed to ON.
ON position
OFF position
For directly-mounted fuse
With board version M or later, the no-fuse breaker (NFB) is replaced with the fuse.
16.10a LAN90B Board
CAUTION
When servicing any printed circuit board, be sure to wear an anti-static wristband to ensure
proper grounding. If not, static electricity built on the human body may cause damage to electric
parts mounted on the board.
LAN90B Board Setting
Make sure that the LAN90B board has been set as shown below.
CN2
CN7
S2
1
009-051-07
08.30.2001 FM3115
13
14
15
16
FR1H1353.EPS
"16.14 How to Check the Fuse on the Board"
S2 setting
2
3
4
5
6
ON
CR-IR341 Service Manual
CN1
CN6
bit
Description
1-3, 5, 6
To be determined
ON: Function test (FT)
4
OFF: OS boot (NORMAL)
MC - 283
S3
Set to "6"
S1
Switch for resetting
the LAN90B board
Setting
Always OFF
Always OFF
FR1B1265.EPS