FujiFilm CR-IR341 Service Manual page 125

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
2.3
Movement (Feed) of Elevation Unit
(1) Overall/General Sequence
In this sequence, the elevation unit is moved to a designated shelf of the cassette set unit in
order to feed IPs.
The movement of the elevation unit cannot be performed if the suction cups are not in their
home position (HP).
If there is any IP being fed, the machine waits until the suction cup movement process
described in "2.6 (3) Sequence of Side-Positioning Conveyance" is completed.
Fig. 2.3.1 Movement of Elevation Unit
Preprocessing: Fig. 2.2.1 and Fig. 3.1.1
N
The suction cup
HP is confirmed.
MB3 moves the
elevation unit to a
designated shelf.
For its parameters,
see Table 2.3.1a.
It is confirmed that
the elevation unit is
in a proper shelf
position.
Table 2.3.1 Operation Parameters
Symbol
Description
Suction cup movement
PB19
(HP
feed leak)
TB17
Wait for next operation
NB101
Elevation movement retry
Table 2.3.1a Elevation Unit Movement Parameters
(PB_)
Movement of
elevation unit
Topmost stage
2nd stage
Topmost stage
3rd stage
Topmost stage
4th stage
2nd stage
topmost stage
2nd stage
3rd stage
2nd stage
4th stage
3rd stage
topmost stage
3rd stage
2nd stage
3rd stage
4th stage
4th stage
topmost stage
4th stage
2nd stage
4th stage
3rd stage
009-051-04
05.31.99 FM2460
3
4
5
6
START
Is the movement of the elevation
unit permitted?
Y
Is SB1 closed?
N
Y
Movement of elevation unit
DRIVE(MB3/PB_)
Is SB3 closed?
N
Y
Retry at the time
END
of shelf position
detection failure
Value
-12 p
0.1 sec
2 times
TR1H0014.EPS
Pulse
Symbol
count (P)
-3,470
PB112
-6,930
PB113
-10,392
PB114
3,470
PB121
-3,460
PB123
-6,922
PB124
6,930
PB131
3,460
PB132
-3,462
PB134
10,392
PB141
6,922
PB142
3,462
PB143
TR1H0015.EPS
CR-IR341 Service Manual
7
8
9
10
It is initiated when the following condition is met.
¥ Movement of the elevation unit permitted: Fig. 2.7.1
The sequence is not initiated until the movement (load) of
(
the elevation unit is performed to load the previous IP and
the loading of that IP is completed.
Level 2 error
Suction cup HP return
23E8
failure
DELAY TB17
Fig.2.3.2
Movement of suction cup
(HP
feed leak)
DRIVE(MB1/PB19)
A
Elevation unit HP positioning
Fig.2.3.3
Is the number of retries made greater
than the retry count (NB101)?
Y
N
Leve 2 error
23DE
Elevation shelf movement
failure retray
A
CW
Cassette feeder
Rack
MB3
4 pulses = 0.1mm travel
(360¡ (= 400 pulses)
Reduction ratio: 1/20
Cassette feeder
Condition as seen from
the cassette loading side
(front)
Retry at the time
of HP detection failure
Level 0 error
03DE
Elevation movement
error
Sensor arm
for SA17
(attached at the
SA17
elevation unit)
Elevation unit
SB3
Sensor arm for SB3
Condition as seen from
the right-hand side
MD - 62
)
FR1H0308.EPS
Elevation
unit
FR1H0309.EPS
MD - 62