Replacement
(1) Remove the shutter assembly.
"4.3 Removing and Reinstalling Shutter Assembly"
(2) Disconnect the connectors (SA1, 5, 9, 13).
(3) Remove SA1, 5, 9, and 13.
The sensor is secured to the bracket by four latches. To remove the sensor, pinch the
latches by fingers.
Cassette set unit
(4) Install a new sensor.
"3.5 Sensors Not Requiring Adjustment of Its Mounting Position" in Machine
Description
(5) Connect the connectors (SA1, 5, 9, 13).
(6) Reinstall the shutter assembly.
"4.3 Removing and Reinstalling Shutter Assembly"
(7) Reinstall the cassette set unit.
"4.2 Removing and Reinstalling Cassette Set Unit"
(8) Check the sensor replaced.
" Checks"
009-051-04
05.31.99 FM2460
3
4
5
6
Bracket
CR-IR341 Service Manual
7
8
9
10
(2)
Connector
(3)
SA1,5,9,13
(3)
Latch
This bracket should
not be removed.
FR1H1055.EPS