FujiFilm CR-IR341 Service Manual page 532

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
Replacement
(1) Remove the cassette set unit.
"4.2 Removing an Reinstalling Cassette Set Unit"
(2) Cut the cable tie.
(3) Disconnect the connectors (CNA4, 8, 12, 16).
(4) Remove the sensor bracket as a whole.
(5) Replace the sensor assembly.
Cassette set unit
SA4
SA8
SA12
SA16
(6) For reinstallation, reverse the removal steps.
When attaching the mounting bracket for SA4, 8, 12, 16, tighten the screw while
holding it down.
CAUTION
When retaining the cables with ties, observe the following precautions.
• Ensure that the cables do not protrude beyond the side plates.
• Retain the cables at proper positions so that they will not be caught by rotating
rollers.
(7) Check the sensor replaced.
" Checks"
009-051-04
05.31.99 FM2460
3
4
5
6
CR-IR341 Service Manual
7
8
9
10
(3)
CNA4, 8, 12, 16
(2)
Cut
cable tie
MC - 41
(4)
Loosen screw
BR M4x8
(5)
Sensor assembly
(4)
Sensor bracket
BR M4x8
FR1H1054.EPS
MC - 41