FujiFilm CR-IR341 Service Manual page 667

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
(10) Select "5. SCANNER UTILITY", "10. VIRTUAL IMAGE", and "4. ROUTINE" in se-
quence.
CAUTION
After checks, be sure to select "ROUTINE" under "VIRTUAL IMAGE" of "SCANNER
UTILITY" to reset the scanner back to its initial condition. If it is not reset to its initial
condition, an image with lower contrast will be outputted.
Checks on high-voltage power supply section
(1) Make sure that the high-voltage switch (S1) of the SCN08A board is in the OFF posi-
tion.
Scanner unit
(2) Remove the light-collecting guide assembly.
"10.2 Servicing Light-Collecting Guide"
(3) Take out the light-collecting guide from the PMT08A board assembly.
PMT08A board assembly
009-051-04
05.31.99 FM2460
3
4
5
6
SCN08A board
Light-collecting guide
CR-IR341 Service Manual
7
8
9
10
High-voltage switch
(1)
OFF position
FR1H1535.EPS
MC - 176
FR1H1800.EPS
MC - 176