FujiFilm CR-IR341 Service Manual page 884

Fuji computed radiography
Table of Contents

Advertisement

C
11
12
ONTENTS
(10) Type in the value of magnetic phase.
The following message appears.
INPUT THE MOVE AND STOP MODE.
1. HI-POWER/ENERGIZED STOP AFTER DELAY
2. LOW-POWER/ENERGIZED STOP AFTER DELAY
3. HI-POWER/DEENERGIZED STOP AFTER DELAY
4. LOW-POWER/DEENERGIZED STOP AFTER DELAY
5. HI-POWER/ENERGIZED STOP
6. LOW-POWER/ENERGIZED STOP
7. HI-POWER/DEENERGIZED STOP
8. LOW-POWER/DEENERGIZED STOP
1 - 8(FIXED=X) :
Meaning: Enter the move and stop mode.
(11) Type in the move/stop mode ("1" through "8").
The following message appears.
INPUT THE STOP MODE.
1. HALT
1 - 2(FIXED=X) :
Meaning: Enter the stop mode.
"X" denotes the stop mode (1: HALT stop, 2: LIMIT stop) that is currently set.
009-051-06
009-051-05
04.20.2001 FM3006
11.20.99 FM2638
13
14
15
16
1: High power driving/energized stop, with power-down delay
2: Low power driving/energized stop, with power-down delay
3: High power driving/deenergized stop, with power-down delay
4: Low power driving/deenergized stop, with power-down delay
5: High power driving/energized stop, without power-down delay
6: Low power driving/energized stop, without power-down delay
7: High power driving/deenergized stop, without power-down delay
8: Low power driving/deenergized stop, without power-down delay
2.LIMIT
CR-IR341 Service Manual