Cleaning and Disinfecting FUJIFILM SonoSite Products User Guide...
Page 2
United States law restricts this device to sale by or on the order of a physician. X-Porte, iViz, Edge, M-Turbo, NanoMaxx, S Series, SonoSite, and the SonoSite logo are registered and unregistered trademarks of FUJIFILM SonoSite, Inc. in various jurisdictions.
Page 3
Chapter 1: Introduction About this guide 1 Conventions 1 Technical support 2 Chapter 2: Cleaning and disinfecting the iViz ultrasound system Cleaning and disinfecting 4 Cleaning and disinfecting the system and protective case 5 Cleaning and disinfecting the transducer 6 Determining the required cleaning and disinfection level 6 Level A: Intermediate-level disinfection (non-critical) 7 Level B: High-level disinfection (semi-critical) 10...
Page 4
Chapter 4: Cleaning and disinfecting the Edge ultrasound system Before getting started 43 Cleaning and disinfecting 44 Determining the required cleaning and disinfecting level 44 Spaulding classifications 44 Clean and disinfect system and transducer to a high level (semi-critical uses) 45 Cleaning and disinfecting the system and transducer (semi-critical)45 Clean and disinfect system and transducer to a low level (non-critical uses) 48 Cleaning and disinfecting the system and transducer (non-critical)49...
Page 5
Determining the required cleaning and disinfecting level 86 Spaulding classifications 86 Clean and disinfect system and transducer to a high level (semi-critical uses) 87 Cleaning and disinfecting the system and transducer (semi-critical)87 Clean and disinfect system and transducer to a low level (non-critical uses) 90 Cleaning and disinfecting the system and transducer (non-critical)91 Storing the transducer 92 Transporting the transducer 93...
Page 6
Cleaning and disinfecting the system and transducer (non-critical)132 Storing the transducer 133 Transporting the transducer 134 Compatible cleaners and disinfectants (system and transducers) 136 Compatible cleaners and disinfectants (system accessories) 139 Compatible high-level disinfectants (transducers) 142...
Page 7
Each system is presented in its own chapter, with cleaning and disinfecting instructions designed specifically for that system. If your FUJIFILM SonoSite system is not listed here, refer to the system user guide, or contact FUJIFILM SonoSite for cleaning instructions.
Page 8
Technical support To order sheaths, bite guards, tip covers, and other supplies, see www.civco.com. For more information or for technical support, contact FUJIFILM SonoSite as follows. Phone (U.S. or Canada) 877-657-8118 Phone (Outside U.S. or 425-951-1330, or call your local representative...
Compatible high-level disinfectants compatible (transducers) ............... 17 This chapter contains instructions for cleaning and disinfecting the iViz ultrasound system, transducers, and accessories. Use the FUJIFILM SonoSite recommendations when cleaning and disinfecting your ultrasound system, transducer, and accessories. Use the cleaning recommendations in the accessory manufacturer’s instructions when cleaning or disinfecting your accessories.
It is important to follow this cleaning and disinfection process without shortening or skipping steps. Only use cleaners and disinfectants approved by FUJIFILM SonoSite on the iViz system. Be sure to observe the solution strength and duration requirements detailed in the cleaning and disinfection procedures.
Do not use strong solvents, such as thinner or benzene, or abrasive cleansers since these will damage the exterior surfaces. Use only FUJIFILM SonoSite recommended cleaners or disinfectants. Use only a soft cloth lightly dampened in a mild soap or compatible cleaning solution to remove any disinfectant on metal surfaces.
Apply the disinfectant solution to the cloth rather than applying it directly to surfaces. Follow the chemical manufacturer’s requirements for wet contact time. Caution Use only FUJIFILM SonoSite recommended disinfectants. Using a non- recommended disinfect solution or incorrect solution strength can damage or discolor the transducer and void the transducer warranty.
Each Spaulding classification mandates a specific level of cleaning and disinfection of the equipment before it can be used in the next exam. Use the following guide to determine the level of cleaning and disinfection for the iViz ultrasound system according to the appropriate Spaulding classification. Table 2-1: Cleaning and disinfection guidelines If any part of the system came into contact Then disinfect to the following level...
Page 14
Table 2-2: Compatible cleaners and disinfectants Protective P21v Product System Carry case case transducer T-Spray II (PI-Spray II) Proper cleaning of the slot is important because good airflow through the slot Note improves cooling of the transducer electronics and extends the life of the equipment.
Page 15
3 Examine the transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, do not use the transducer and contact FUJIFILM SonoSite or your local representative. Determining the required cleaning and disinfection level...
Level B: High-level disinfection (semi-critical) Clean and disinfect the transducer immediately after use. Be sure to follow the manufacturer's instructions when using cleaners and disinfectants. Refer to Table 2-3, “Compatible cleaners and disinfectants”on page 10 for a list of cleaners and disinfectants recommended for use with the transducer.
Page 17
17. Caution Use only FUJIFILM SonoSite recommended disinfectants. Using a non- recommended disinfection solution, incorrect solution strength, or immersing a transducer deeper or for a longer period of time than recommended can damage or discolor the transducer and void the transducer warranty.
Page 18
4 Examine the transducer and cord for damage, such as cracks or splitting where fluid can enter. If damage is evident, discontinue use of the transducer and contact FUJIFILM SonoSite or your local representative...
If the iViz ultrasound system shows damage or deterioration, such as corrosion, discoloration, pitting, or cracking, do not dispose of it. Contact FUJIFILM SonoSite or your local representative. The battery inside this unit may explode if exposed to very high temperatures. Do not destroy this unit by incinerating or burning.
Compatible cleaners and disinfectants (system and transducers) Table 2-5: Cleaners and disinfectants compatible with the iViz system and transducers Disinfectants Airkem A-33 Alcohol-Ethanol Aniosyme DD1- 5% Anioxide 1000 Asepti Wipe II Baxedin Pre-Op Caviwipes Cidalkan Cidalkan Lingettes Clinell Universal Sanitising Wipes Control III Coverage Germicidal Surface...
Page 21
Table 2-5: Cleaners and disinfectants compatible with the iViz system and transducers Disinfectants Dispatch Wipes Dynacide PA Echo Clean lingettes End Bac II Endosporine Envirocide Gluco Chlor Wipes Hydrogen Peroxide 3% Incidin Plus 1% Incidin Plus 3% Korsolex Extra Medzyme MetriSponge Percept RTU Wipes Protex Spray...
Page 22
Table 2-5: Cleaners and disinfectants compatible with the iViz system and transducers Disinfectants Sani-Cloth Bleach Wipes Sani-Cloth HB Sani-Cloth Plus Staphene Steranios 2% Super Sanicloth Trigene Advance Wipes Tristel Solo Tristel Wipes T-Spray T-Spray II Virex TB Wet Task Wipes Wex-Cide Wipes Plus Disinfecting Wipes Compatible cleaners and disinfectants (system and transducers)
Page 25
Compatible high-level disinfectants (transducers) ................... 38 This chapter contains instructions for cleaning and disinfecting the X-Porte ultrasound system, transducers, and accessories. Use the FUJIFILM SonoSite recommendations when cleaning or disinfecting your ultrasound system, transducer, and accessories. Use the cleaning recommendations in the component manufacturer’s instructions when cleaning or disinfecting your components.
(PPE), such as protective eyewear and gloves. Inspect the system to determine that it is free of any unacceptable deterioration, such as corrosion, discoloration, pitting, or cracked seals. If damage is evident, discontinue use, and contact FUJIFILM SonoSite or your local representative. ...
Cleaning and disinfecting Determining the required cleaning and disinfecting level The cleaning instructions contained in this chapter are based on requirements mandated by the American Food and Drug Administration (FDA). Failure to follow WARNINGS these instructions may result in cross contamination and patient infection. The level of cleaning and disinfecting required for the system is dictated by the type of tissue it contacts during use.
This can damage the transducer and void the warranty. Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
Page 29
5 Clean the exterior surfaces of the ULTRASOUND SYSTEM to remove any debris or bodily fluids. Use the following procedure: a Use either a premoistened wipe or a soft cloth dampened with cleaner or disinfectant. Choose a cleaner from Table 3-2, “Cleaners and disinfectants compatible with X-Porte system and transducers” page 30.
Page 30
Do not immerse the transducer connector in any disinfectant solution. Cautions Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage or discolor the transducer and void the warranty.
12 Dispose of the disinfectant according to the manufacturer’s guidelines. 13 Examine the transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, discontinue use of the transducer, and contact FUJIFILM SonoSite or your local representative.
Page 32
Doing so may cause solution to leak into the system, damaging it and voiding the warranty. Use only FUJIFILM SonoSite-recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can Cautions damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
Allow the ultrasound system, including the console and transducer, to air dry in a clean, well-ventilated space. Examine the system, transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, do not use the transducer. Instead, contact FUJIFILM SonoSite or your local representative. Storing the transducer To store the transducer 1 Make sure the transducer has been cleaned and disinfected as detailed in this chapter.
Store the transducer in an environment that is safe and has good airflow. Do not store the transducer in closed containers or where condensation may occur. Avoid direct sunlight and exposure to x-rays. Recommended storage temperature range is between 0 F) and +45 C (113 If using a wall-mounted rack for storage, ensure that:...
Page 35
WARNINGS this chapter or according to special instructions received from FUJIFILM SonoSite. If you are returning the transducer to FUJIFILM SonoSite, document the disinfection on a “Declaration of Cleanliness, ” and attach it to the packing list.
Compatible cleaners and disinfectants (system and transducers) Table 3-2: Cleaners and disinfectants compatible with X-Porte system and transducers Disinfectants Airkem A-33 Alcohol-Ethanol Anios Wipes Aniosyme DD1- 5% Anioxide 1000 Asepti HB Asepti Wipe II Baxedin Pre-Op Carpe Diem TB Wipes Caviwipes Cidalkan Lingettes Clinell Universal...
Page 37
Table 3-2: Cleaners and disinfectants compatible with X-Porte system and transducers Disinfectants Control III Coverage Germicidal Surface Wipes Coverage Plus Germicidal Surface Wipes Coverage Spray Coverage Spray TB Dynacide PA Echo Clean lingettes End Bac II Endosporine Envirocide Gluco Chlor Wipes Hydrogen Peroxide Incidin Plus 1% Incidin Plus 3%...
Page 38
Table 3-2: Cleaners and disinfectants compatible with X-Porte system and transducers Disinfectants MetriSponge Metrizyme Mikrobac Forte 2% Mikrobac Tissues Oxivir TB Wipes Percept RTU Protex Spray Protex Wipes Rely+On Perasafe Sani-Cloth AF Sani-Cloth AF3 Sani-Cloth Bleach Wipes Sani-Cloth HB Sani-Cloth Plus Simple Chlor Staphene Compatible cleaners and...
Page 39
Table 3-2: Cleaners and disinfectants compatible with X-Porte system and transducers Disinfectants Steranios 2% Steranios 20% Trigene Advance Wipes Tristel Solo T-Spray T-Spray II Virex II 256 Virex TB Virox 5 RTU Wipes Wet Task Wipes Wex-Cide Wipes Plus Disinfecting Wipes Wood Wyant Ultra Wipes Compatible cleaners and disinfectants (system and transducers)
Page 47
Compatible high-level disinfectants (transducers) ................... 61 This chapter contains instructions for cleaning and disinfecting the Edge ultrasound system, transducers, and accessories. Use the FUJIFILM SonoSite recommendations when cleaning or disinfecting your ultrasound system, transducer, and accessories. Use the cleaning recommendations in the component manufacturer’s instructions when cleaning or disinfecting your components.
Page 48
Ensure that cleaning and disinfecting solutions and wipes are not expired. WARNINGS Some cleaners and disinfectants can cause an allergic reaction to some individuals.
(PPE), such as protective eyewear and gloves. Inspect the system to determine that it is free of any unacceptable deterioration, such as corrosion, discoloration, pitting, or cracked seals. If damage is evident, discontinue use, and contact FUJIFILM SonoSite or your local representative. ...
Cleaning and disinfecting Determining the required cleaning and disinfecting level The cleaning instructions contained in this chapter are based on requirements mandated by the American Food and Drug Administration (FDA). Failure to follow WARNINGS these instructions may result in cross contamination and patient infection. The level of cleaning and disinfecting required for the system is dictated by the type of tissue it contacts during use.
This can damage the transducer and void the warranty. Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
Page 52
5 Clean the exterior surfaces of the ULTRASOUND SYSTEM to remove any debris or bodily fluids. Use the following procedure: a Use either a premoistened wipe or a soft cloth dampened with cleaner or disinfectant. Choose a cleaner from Table 4-2, “Cleaners and disinfectants compatible with Edge system and transducers”on page 53.
Page 53
Do not immerse the transducer connector in any disinfectant solution. Cautions Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage or discolor the transducer and void the warranty.
12 Dispose of the disinfectant according to the manufacturer’s guidelines. 13 Examine the transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, discontinue use of the transducer, and contact FUJIFILM SonoSite or your local representative.
Page 55
Doing so may cause solution to leak into the system, damaging it and voiding the warranty. Use only FUJIFILM SonoSite-recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can Cautions damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
Allow the ultrasound system, including the console and transducer, to air dry in a clean, well-ventilated space. Examine the system, transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, do not use the transducer. Instead, contact FUJIFILM SonoSite or your local representative. Storing the transducer To store the transducer 1 Make sure the transducer has been cleaned and disinfected as detailed in this chapter.
Store the transducer in an environment that is safe and has good airflow. Do not store the transducer in closed containers or where condensation may occur. Avoid direct sunlight and exposure to x-rays. Recommended storage temperature range is between 0 F) and +45 C (113 If using a wall-mounted rack for storage, ensure that:...
Page 58
WARNINGS this chapter or according to special instructions received from FUJIFILM SonoSite. If you are returning the transducer to FUJIFILM SonoSite, document the disinfection on a “Declaration of Cleanliness, ” and attach it to the packing list.
Compatible cleaners and disinfectants (system and transducers Table 4-2: Cleaners and disinfectants compatible with Edge system and transducers Disinfectants Accel Intervention Wipes Accel TB Wipes Airkem A-33 Alcohol-Ethanol Alcohol - Isopropyl Anios Wipes Aniosyme DD1- 5% Anioxide 1000 Asepti HB Asepti Wipe II Baxedin Pre-Op Caviwipes...
Page 60
Table 4-2: Cleaners and disinfectants compatible with Edge system and transducers Disinfectants Coverage Germicidal Surface Wipes Coverage Plus Germicidal Surface Wipes Coverage Spray Coverage Spray TB Dynacide PA Echo Clean lingettes End Bac II Endosporine Envirocide Gluco Chlor Wipes Hydrogen Peroxide Incidin Plus 1% Incidin Plus 3% Medzyme...
Page 61
Table 4-2: Cleaners and disinfectants compatible with Edge system and transducers Disinfectants Mikrobac Forte 2% Mikrobac Tissues Oxivir TB Wipes Percept RTU Wipes Protex Spray Protex Wipes Rely+On Perasafe Sani-Cloth AF Sani-Cloth AF3 Sani-Cloth Bleach Wipes Sani-Cloth HB Sani-Cloth Plus Staphene Steranios 2% Steranios 20%...
Table 4-2: Cleaners and disinfectants compatible with Edge system and transducers Disinfectants Tristel Wipes T-Spray T-Spray II Virex II 256 Virex TB Virox 5 RTU Wipes Wet Task Wipes Wex-Cide Wipes Plus Disinfecting Wipes Compatible cleaners and disinfectants (system accessories) Table 4-3: Cleaners and disinfectants compatible with Edge system accessories Disinfectants Accel Intervention Wipes...
Page 63
Table 4-3: Cleaners and disinfectants compatible with Edge system accessories Disinfectants Airkem A-33 Alcohol-Ethanol Alcohol - Isopropyl Anios Wipes Aniosyme DD1- 5% Anioxide 1000 Asepti HB Asepti Steryl Asepti Wipe II Baxedin Pre-Op Carpe Diem TB Wipes Caviwipes Cidalkan Cidalkan Lingettes Clinell Universal Sanitising Wipes Clorox Germicidal Wipes...
Page 64
Table 4-3: Cleaners and disinfectants compatible with Edge system accessories Disinfectants Clorox Hydrogen Peroxide Wipes Control III Coverage Germicidal Surface Wipes Coverage Plus Germicidal Surface Wipes Coverage Spray Coverage Spray TB DisCide Ultra Disinfecting Towlettes Dynacide PA Echo Clean lingettes End Bac II Endosporine Envirocide...
Page 65
Table 4-3: Cleaners and disinfectants compatible with Edge system accessories Disinfectants Medzyme Mikrobac Forte 2% Mikrobac Tissues Mikrozid Tucher NU - Cidex Oxivir TB Wipes Percept RTU Wipes Protex Spray Protex Wipes Rely+On Perasafe Sani-Cloth AF Sani-Cloth AF3 Sani-Cloth Bleach Wipes Sani-Cloth HB Sani-Cloth Plus Simple Chlor...
Page 66
Table 4-3: Cleaners and disinfectants compatible with Edge system accessories Disinfectants Steranios 2% Steranios 20% Super Sani-Cloth Trigene Advance Wipes Tristel Solo Tristel Wipes T-Spray T-Spray II Virex II 256 Virex TB Virox 5 RTU Wipes Wet Task Wipes Wex-Cide Wipes Plus Disinfecting Wipes Wood Wyant Ultra Wipes Compatible cleaners and disinfectants...
Page 69
Also included is a listing of disinfectants that FUJIFILM SonoSite has tested and deemed compatible with the SonoSite Edge II system, transducers, and accessories. This information is presented in the following tables. ...
(PPE), such as protective eyewear and gloves. Inspect the system to determine that it is free of any unacceptable deterioration, such as corrosion, discoloration, pitting, or cracked seals. If damage is evident, discontinue use, and contact FUJIFILM SonoSite or your local representative. ...
Cleaning and disinfecting Determining the required cleaning and disinfecting level The cleaning instructions contained in this chapter are based on requirements mandated by the American Food and Drug Administration (FDA). Failure to follow WARNING these instructions may result in cross contamination and patient infection. The level of cleaning and disinfecting required for the system is dictated by the type of tissue it contacts during use.
This can damage the transducer and void the warranty. Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
Page 73
5 Clean the exterior surfaces of the ULTRASOUND SYSTEM to remove any debris or bodily fluids. Use the following procedure: a Use either a premoistened wipe or a soft cloth dampened with cleaner or disinfectant. Choose a cleaner from Table 5-2, “Cleaners and disinfectants compatible with SonoSite Edge II system and transducers”on page 74.
Page 74
Do not immerse the transducer connector in any disinfectant solution. Cautions Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage or discolor the transducer and void the warranty.
12 Dispose of the disinfectant according to the manufacturer’s guidelines. 13 Examine the transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, discontinue use of the transducer, and contact FUJIFILM SonoSite or your local representative.
Page 76
Doing so may cause solution to leak into the system, damaging it and voiding the warranty. Use only FUJIFILM SonoSite-recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can Cautions damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
Allow the ultrasound system, including the console and transducer, to air dry in a clean, well-ventilated space. Examine the system, transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, do not use the transducer. Instead, contact FUJIFILM SonoSite or your local representative. Storing the transducer To store the transducer 1 Make sure the transducer has been cleaned and disinfected as detailed in this chapter.
Store the transducer in an environment that is safe and has good airflow. Do not store the transducer in closed containers or where condensation may occur. Avoid direct sunlight and exposure to x-rays. Recommended storage temperature range is between 0 F) and +45 C (113 If using a wall-mounted rack for storage, ensure that:...
Page 79
WARNING this chapter or according to special instructions received from FUJIFILM SonoSite. If you are returning the transducer to FUJIFILM SonoSite, document the disinfection on a “Declaration of Cleanliness, ” and attach it to the packing list.
Compatible cleaners and disinfectants (system and transducers) Table 5-2: Cleaners and disinfectants compatible with SonoSite Edge II system and transducers Disinfectants Accel Intervention Wipes Accel TB Wipes Airkem A-33 Alcohol-Ethanol Alcohol - Isopropyl Anios Wipes Aniosyme DD1- 5% Anioxide 1000 Asepti HB Asepti Steryl Asepti Wipe II...
Page 81
Table 5-2: Cleaners and disinfectants compatible with SonoSite Edge II system and transducers Disinfectants Clorox Germicidal Wipes Clorox Hydrogen Peroxide Wipes Control III Coverage Germicidal Surface Wipes Coverage Plus Germicidal Surface Wipes Coverage Spray Coverage Spray TB DisCide Ultra Disinfecting Towlettes Dynacide PA Echo Clean lingettes End Bac II...
Page 82
Table 5-2: Cleaners and disinfectants compatible with SonoSite Edge II system and transducers Disinfectants Incidin Plus 3% Medzyme MetriSponge Mikrobac Forte 2% Mikrobac Tissues Oxivir TB Wipes Percept RTU Wipes Protex Spray Protex Wipes Rely+On Perasafe Sani-Cloth AF Sani-Cloth AF3 Sani-Cloth Bleach Wipes Sani-Cloth HB...
Page 83
Table 5-2: Cleaners and disinfectants compatible with SonoSite Edge II system and transducers Disinfectants Trigene Advance Wipes Tristel Solo Tristel Wipes T-Spray T-Spray II Virex II 256 Virex TB Virox 5 RTU Wipes Wet Task Wipes Wex-Cide Wipes Plus Disinfecting Wipes Wood Wyant Ultra Wipes Compatible cleaners and disinfectants (system and transducers)
Compatible cleaners and disinfectants (system accessories) Table 5-3: Cleaners and disinfectants compatible with SonoSite Edge II system accessories Disinfectants Accel Intervention Wipes Accel TB Wipes Airkem A-33 Alcohol-Ethanol Alcohol - Isopropyl Anios Wipes Aniosyme DD1- 5% Anioxide 1000 Anioxy-Twin Asepti HB Asepti Steryl Asepti Wipe II Baxedin Pre-Op...
Page 85
Table 5-3: Cleaners and disinfectants compatible with SonoSite Edge II system accessories Disinfectants Control III Coverage Spray Coverage Spray TB DisCide Ultra Disinfecting Towlettes Dispatch Wipes Dynacide PA Echo Clean lingettes End Bac II Endosporine Envirocide Gluco Chlor Wipes Hydrogen Peroxide 3% Incidin Plus 1% Incidin Plus 3% Medzyme...
Page 86
Table 5-3: Cleaners and disinfectants compatible with SonoSite Edge II system accessories Disinfectants Rely+On Perasafe Sani-Cloth AF Sani-Cloth AF3 Sani-Cloth Bleach Wipes Sani-Cloth HB Sani-Cloth Plus Staphene Steranios 2% Steranios 20% Super Sani-Cloth Trigene Advance Wipes Tristel Solo Tristel Wipes T-Spray T-Spray II Virex II 256...
Page 89
Compatible high-level disinfectants (transducers) ................. 103 This chapter contains instructions for cleaning and disinfecting the S Series ultrasound system, transducers and accessories. Use the FUJIFILM SonoSite recommendations when cleaning or disinfecting your ultrasound system, transducer, and accessories. Use the cleaning recommendations in the component manufacturer’s instructions when cleaning or disinfecting your components.
Page 90
Ensure that cleaning and disinfecting solutions and wipes are not expired WARNINGS Some cleaners and disinfectants can cause an allergic reaction to some individuals.
(PPE), such as protective eyewear and gloves. Inspect the system to determine that it is free of any unacceptable deterioration, such as corrosion, discoloration, pitting, or cracked seals. If damage is evident, discontinue use, and contact FUJIFILM SonoSite or your local representative. ...
Cleaning and disinfecting Determining the required cleaning and disinfecting level The cleaning instructions contained in this chapter are based on requirements mandated by the American Food and Drug Administration (FDA). Failure to follow WARNINGS these instructions may result in cross contamination and patient infection. The level of cleaning and disinfecting required for the system is dictated by the type of tissue it contacts during use.
This can damage the transducer and void the warranty. Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
Page 94
5 Clean the exterior surfaces of the ULTRASOUND SYSTEM to remove any debris or bodily fluids. Use the following procedure: a Use either a premoistened wipe or a soft cloth dampened with cleaner or disinfectant. Choose a cleaner from Table 6-2, “Cleaners and disinfectants compatible with S Series system and transducers” page 95.
Page 95
Do not immerse the transducer connector in any disinfectant solution. Cautions Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage or discolor the transducer and void the warranty.
12 Dispose of the disinfectant according to the manufacturer’s guidelines. 13 Examine the transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, discontinue use of the transducer, and contact FUJIFILM SonoSite or your local representative.
Page 97
Doing so may cause solution to leak into the system, damaging it and voiding the warranty. Use only FUJIFILM SonoSite-recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can Cautions damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
Allow the ultrasound system, including the console and transducer, to air dry in a clean, well-ventilated space. Examine the system, transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, do not use the transducer. Instead, contact FUJIFILM SonoSite or your local representative. Storing the transducer To store the transducer 1 Make sure the transducer has been cleaned and disinfected as detailed in this chapter.
Store the transducer in an environment that is safe and has good airflow. Do not store the transducer in closed containers or where condensation may occur. Avoid direct sunlight and exposure to x-rays. Recommended storage temperature range is between 0 F) and +45 C (113 If using a wall-mounted rack for storage, ensure that:...
Page 100
WARNINGS this chapter or according to special instructions received from FUJIFILM SonoSite. If you are returning the transducer to FUJIFILM SonoSite, document the disinfection on a “Declaration of Cleanliness, ” and attach it to the packing list.
Compatible cleaners and disinfectants (system and transducers) Table 6-2: Cleaners and disinfectants compatible with S Series system and transducers Disinfectants Accel TB Wipes Airkem A-33 Alcohol-Ethanol Alcohol - Isopropyl Anios Wipes Aniosyme DD1- 5% Asepti HB Asepti Wipe II Bacoban WB Baxedin Pre-Op Carpe Diem TB Wipes...
Page 102
Table 6-2: Cleaners and disinfectants compatible with S Series system and transducers Disinfectants Clorox Germicidal Wipes Clorox Hydrogen Peroxide Wipes Control III Coverage Germicidal Surface Wipes Coverage Plus Germicidal Surface Wipes Coverage Spray Coverage Spray TB DisCide Ultra Disinfecting Towlettes Dynacide PA Echo Clean lingettes...
Page 103
Table 6-2: Cleaners and disinfectants compatible with S Series system and transducers Disinfectants Hydrogen Peroxide Incidin Plus 1% Incidin Plus 3% Medzyme MetriSponge Mikrobac Forte 2% Mikrobac Tissues Oxivir TB Wipes Percept RTU Wipes Protex Spray Protex Wipes Sani-Cloth AF Sani-Cloth AF3 Sani-Cloth Bleach Wipes...
Page 104
Table 6-2: Cleaners and disinfectants compatible with S Series system and transducers Disinfectants Simple Chlor Staphene Steranios 2% Steranios 20% Trigene Advance Wipes Tristel Solo Tristel Wipes T-Spray T-Spray II Virex II 256 Virex TB Virox 5 RTU Wipes Wet Task Wipes Wex-Cide Wipes Plus Disinfecting Wipes...
Compatible cleaners and disinfectants (system accessories) Table 6-3: Cleaners and disinfectants compatible with S Series system accessories Disinfectants Accel Intervention Wipes Accel TB Wipes Airkem A-33 Alcohol-Ethanol Alcohol - Isopropyl Anios Wipes Aniosyme DD1- 5% Anioxide 1000 Asepti HB Asepti Steryl Asepti Wipe II Baxedin Pre-Op Carpe Diem TB Wipes...
Page 106
Table 6-3: Cleaners and disinfectants compatible with S Series system accessories Disinfectants Clorox Germicidal Wipes Clorox Hydrogen Peroxide Wipes Control III Coverage Germicidal Surface Wipes Coverage Plus Germicidal Surface Wipes Coverage Spray Coverage Spray TB DisCide Ultra Disinfecting Towlettes Dynacide PA Echo Clean lingettes End Bac II Endosporine...
Page 107
Table 6-3: Cleaners and disinfectants compatible with S Series system accessories Disinfectants Incidin Plus 3% Medzyme Metrizyme Mikrobac Forte 2% Mikrobac Tissues NU - Cidex Oxivir TB Wipes Percept RTU Wipes Protex Spray Protex Wipes Rely+On Perasafe Sani-Cloth AF Sani-Cloth AF3 Sani-Cloth Bleach Wipes Sani-Cloth HB Sani-Cloth Plus...
Page 108
Table 6-3: Cleaners and disinfectants compatible with S Series system accessories Disinfectants Staphene Steranios 2% Steranios 20% Super Sani-Cloth Trigene Advance Wipes Tristel Solo Tristel Wipes T-Spray T-Spray II Virex II 256 Virex TB Virox 5 RTU Wipes Wet Task Wipes Wex-Cide Wipes Plus Disinfecting Wipes Wood Wyant Ultra Wipes...
Page 111
Compatible high-level disinfectants (transducers) ................. 124 This chapter contains instructions for cleaning and disinfecting the M-Turbo ultrasound system, transducers, and accessories. Use the FUJIFILM SonoSite recommendations when cleaning or disinfecting your ultrasound system, transducer, and accessories. Use the cleaning recommendations in the component manufacturer’s instructions when cleaning or disinfecting your components.
(PPE), such as protective eyewear and gloves. Inspect the system to determine that it is free of any unacceptable deterioration, such as corrosion, discoloration, pitting, or cracked seals. If damage is evident, discontinue use, and contact FUJIFILM SonoSite or your local representative. ...
Cleaning and disinfecting Determining the required cleaning and disinfecting level The cleaning instructions contained in this chapter are based on requirements mandated by the American Food and Drug Administration (FDA). Failure to follow WARNINGS these instructions may result in cross contamination and patient infection. The level of cleaning and disinfecting required for the system is dictated by the type of tissue it contacts during use.
This can damage the transducer and void the warranty. Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
Page 115
5 Clean the exterior surfaces of the ULTRASOUND SYSTEM to remove any debris or bodily fluids. Use the following procedure: a Use either a premoistened wipe or a soft cloth dampened with cleaner or disinfectant. Choose a cleaner from Table 7-2, “Cleaners and disinfectants compatible with M-Turbo system and transducers”on page 116.
Page 116
Do not immerse the transducer connector in any disinfectant solution. Cautions Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage or discolor the transducer and void the warranty.
12 Dispose of the disinfectant according to the manufacturer’s guidelines. 13 Examine the transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, discontinue use of the transducer, and contact FUJIFILM SonoSite or your local representative.
Page 118
Doing so may cause solution to leak into the system, damaging it and voiding the warranty. Use only FUJIFILM SonoSite-recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can Cautions damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
Allow the ultrasound system, including the console and transducer, to air dry in a clean, well-ventilated space. Examine the system, transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, do not use the transducer. Instead, contact FUJIFILM SonoSite or your local representative. Storing the transducer To store the transducer 1 Make sure the transducer has been cleaned and disinfected as detailed in this chapter.
Store the transducer in an environment that is safe and has good airflow. Do not store the transducer in closed containers or where condensation may occur. Avoid direct sunlight and exposure to x-rays. Recommended storage temperature range is between 0 F) and +45 C (113 If using a wall-mounted rack for storage, ensure that:...
Page 121
WARNINGS this chapter or according to special instructions received from FUJIFILM SonoSite. If you are returning the transducer to FUJIFILM SonoSite, document the disinfection on a “Declaration of Cleanliness, ” and attach it to the packing list.
Compatible cleaners and disinfectants (system and transducers) Table 7-2: Cleaners and disinfectants compatible with M-Turbo system and transducers Disinfectants Accel Intervention Wipes Accel Prevention Airkem A-33 Alcohol-Ethanol Alcohol - Isopropyl Anios Wipes Aniosyme DD1- Anioxide 1000 Asepti HB Asepti Wipe II Baxedin Pre-Op Carpe Diem TB Wipes...
Page 123
Table 7-2: Cleaners and disinfectants compatible with M-Turbo system and transducers Disinfectants Clorox Germicidal Wipes Clorox Hydrogen Peroxide Wipes Control III Coverage Germicidal Surface Wipes Coverage Plus Germicidal Surface Wipes Coverage Spray Coverage Spray TB Dynacide PA Echo Clean lingettes End Bac II Endosporine Envirocide...
Page 124
Table 7-2: Cleaners and disinfectants compatible with M-Turbo system and transducers Disinfectants Incidin Plus 1% Incidin Plus 3% Medzyme MetriSponge Mikrobac Forte 2% Mikrobac Tissues Oxivir TB Wipes Percept RTU Wipes Protex Spray Protex Wipes Sani-Cloth AF Sani-Cloth AF3 Sani-Cloth Bleach Wipes Sani-Cloth HB Sani-Cloth Plus...
Page 125
Table 7-2: Cleaners and disinfectants compatible with M-Turbo system and transducers Disinfectants Steranios 2% Steranios 20% Trigene Advance Wipes Tristel Solo Tristel Wipes T-Spray T-Spray II Virex II 256 Virex TB Virox 5 RTU Wipes Wet Task Wipes Wex-Cide Wipes Plus Disinfecting Wipes Compatible cleaners and disinfectants (system and transducers)
Compatible high-level disinfectants (transducers) Table 7-4: High-level disinfectants compatible with M-Turbo transducers High-level Disinfectants Cidex Cidex OPA Metricide OPA Plus Madacide-1 Revital-Ox Resert HLD Cidex Plus Compatible high-level disinfectants (transducers)
Page 131
Compatible high-level disinfectants (transducers) ................. 142 This chapter contains instructions for cleaning and disinfecting the NanoMaxx ultrasound system, transducers, and accessories. Use the FUJIFILM SonoSite recommendations when cleaning or disincomponentfecting your ultrasound system, transducer, and accessories. Use the cleaning recommendations in the component manufacturer’s instructions when cleaning or disinfecting your components.
(PPE), such as protective eyewear and gloves. Inspect the system to determine that it is free of any unacceptable deterioration, such as corrosion, discoloration, pitting, or cracked seals. If damage is evident, discontinue use, and contact FUJIFILM SonoSite or your local representative. ...
Cleaning and disinfecting Determining the required cleaning and disinfecting level The cleaning instructions contained in this chapter are based on requirements mandated by the American Food and Drug Administration (FDA). Failure to follow WARNING these instructions may result in cross contamination and patient infection. The level of cleaning and disinfecting required for the system is dictated by the type of tissue it contacts during use.
This can damage the transducer and void the warranty. Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
Page 135
5 Clean the exterior surfaces of the ULTRASOUND CONSOLE to remove any debris or bodily fluids. Use the following procedure: a Use either a premoistened wipe or a soft cloth dampened with cleaner or disinfectant. Choose a cleaner from Table 8-2, “Cleaners and disinfectants compatible with NanoMaxx system and transducers”on page 136.
Page 136
Do not immerse the transducer connector in any disinfectant solution. Cautions Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage or discolor the transducer and void the warranty.
12 Dispose of the disinfectant according to the manufacturer’s guidelines. 13 Examine the transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, discontinue use of the transducer, and contact FUJIFILM SonoSite or your local representative.
Page 138
Doing so may cause solution to leak into the system, damaging it and voiding the warranty. Use only FUJIFILM SonoSite-recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can Cautions damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
Allow the ultrasound system, including the console and transducer, to air dry in a clean, well-ventilated space. Examine the system, transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, do not use the transducer. Instead, contact FUJIFILM SonoSite or your local representative. Storing the transducer To store the transducer 1 Make sure the transducer has been cleaned and disinfected as detailed in this chapter.
Store the transducer in an environment that is safe and has good airflow. Do not store the transducer in closed containers or where condensation may occur. Avoid direct sunlight and exposure to x-rays. Recommended storage temperature range is between 0 F) and +45 C (113 If using a wall-mounted rack for storage, ensure that:...
Page 141
WARNING this chapter or according to special instructions received from FUJIFILM SonoSite. If you are returning the transducer to FUJIFILM SonoSite, document the disinfection on a “Declaration of Cleanliness, ” and attach it to the packing list.