Cleaning and Disinfecting
FUJIFILM SonoSite Products
User Guide

Advertisement

loading
Need help?

Need help?

Do you have a question about the SonoSite S Series and is the answer not in the manual?

Questions and answers

Summary of Contents for FujiFilm SonoSite S Series

  • Page 1 Cleaning and Disinfecting FUJIFILM SonoSite Products User Guide...
  • Page 2 United States law restricts this device to sale by or on the order of a physician. X-Porte, iViz, Edge, M-Turbo, NanoMaxx, S Series, SonoSite, and the SonoSite logo are registered and unregistered trademarks of FUJIFILM SonoSite, Inc. in various jurisdictions.
  • Page 3 Chapter 1: Introduction About this guide 1 Conventions 1 Technical support 2 Chapter 2: Cleaning and disinfecting the iViz ultrasound system Cleaning and disinfecting 4 Cleaning and disinfecting the system and protective case 5 Cleaning and disinfecting the transducer 6 Determining the required cleaning and disinfection level 6 Level A: Intermediate-level disinfection (non-critical) 7 Level B: High-level disinfection (semi-critical) 10...
  • Page 4 Chapter 4: Cleaning and disinfecting the Edge ultrasound system Before getting started 43 Cleaning and disinfecting 44 Determining the required cleaning and disinfecting level 44 Spaulding classifications 44 Clean and disinfect system and transducer to a high level (semi-critical uses) 45 Cleaning and disinfecting the system and transducer (semi-critical)45 Clean and disinfect system and transducer to a low level (non-critical uses) 48 Cleaning and disinfecting the system and transducer (non-critical)49...
  • Page 5 Determining the required cleaning and disinfecting level 86 Spaulding classifications 86 Clean and disinfect system and transducer to a high level (semi-critical uses) 87 Cleaning and disinfecting the system and transducer (semi-critical)87 Clean and disinfect system and transducer to a low level (non-critical uses) 90 Cleaning and disinfecting the system and transducer (non-critical)91 Storing the transducer 92 Transporting the transducer 93...
  • Page 6 Cleaning and disinfecting the system and transducer (non-critical)132 Storing the transducer 133 Transporting the transducer 134 Compatible cleaners and disinfectants (system and transducers) 136 Compatible cleaners and disinfectants (system accessories) 139 Compatible high-level disinfectants (transducers) 142...
  • Page 7 Each system is presented in its own chapter, with cleaning and disinfecting instructions designed specifically for that system. If your FUJIFILM SonoSite system is not listed here, refer to the system user guide, or contact FUJIFILM SonoSite for cleaning instructions.
  • Page 8 Technical support To order sheaths, bite guards, tip covers, and other supplies, see www.civco.com. For more information or for technical support, contact FUJIFILM SonoSite as follows. Phone (U.S. or Canada) 877-657-8118 Phone (Outside U.S. or 425-951-1330, or call your local representative...
  • Page 9: Table Of Contents

    Compatible high-level disinfectants compatible (transducers) ............... 17 This chapter contains instructions for cleaning and disinfecting the iViz ultrasound system, transducers, and accessories. Use the FUJIFILM SonoSite recommendations when cleaning and disinfecting your ultrasound system, transducer, and accessories. Use the cleaning recommendations in the accessory manufacturer’s instructions when cleaning or disinfecting your accessories.
  • Page 10: Cleaning And Disinfecting

    It is important to follow this cleaning and disinfection process without shortening or skipping steps. Only use cleaners and disinfectants approved by FUJIFILM SonoSite on the iViz system. Be sure to observe the solution strength and duration requirements detailed in the cleaning and disinfection procedures.
  • Page 11: Cleaning And Disinfecting The System And Protective Case

    Do not use strong solvents, such as thinner or benzene, or abrasive cleansers since these will damage the exterior surfaces. Use only FUJIFILM SonoSite recommended cleaners or disinfectants. Use only a soft cloth lightly dampened in a mild soap or compatible cleaning solution to remove any disinfectant on metal surfaces.
  • Page 12: Cleaning And Disinfecting The Transducer

    Apply the disinfectant solution to the cloth rather than applying it directly to surfaces. Follow the chemical manufacturer’s requirements for wet contact time. Caution Use only FUJIFILM SonoSite recommended disinfectants. Using a non- recommended disinfect solution or incorrect solution strength can damage or discolor the transducer and void the transducer warranty.
  • Page 13: Level A: Intermediate-Level Disinfection (Non-Critical)

    Each Spaulding classification mandates a specific level of cleaning and disinfection of the equipment before it can be used in the next exam. Use the following guide to determine the level of cleaning and disinfection for the iViz ultrasound system according to the appropriate Spaulding classification. Table 2-1: Cleaning and disinfection guidelines If any part of the system came into contact Then disinfect to the following level...
  • Page 14 Table 2-2: Compatible cleaners and disinfectants Protective P21v Product System Carry case case transducer T-Spray II (PI-Spray II) Proper cleaning of the slot is important because good airflow through the slot Note improves cooling of the transducer electronics and extends the life of the equipment.
  • Page 15 3 Examine the transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, do not use the transducer and contact FUJIFILM SonoSite or your local representative. Determining the required cleaning and disinfection level...
  • Page 16: Level B: High-Level Disinfection (Semi-Critical)

    Level B: High-level disinfection (semi-critical) Clean and disinfect the transducer immediately after use. Be sure to follow the manufacturer's instructions when using cleaners and disinfectants. Refer to Table 2-3, “Compatible cleaners and disinfectants”on page 10 for a list of cleaners and disinfectants recommended for use with the transducer.
  • Page 17 17. Caution Use only FUJIFILM SonoSite recommended disinfectants. Using a non- recommended disinfection solution, incorrect solution strength, or immersing a transducer deeper or for a longer period of time than recommended can damage or discolor the transducer and void the transducer warranty.
  • Page 18 4 Examine the transducer and cord for damage, such as cracks or splitting where fluid can enter. If damage is evident, discontinue use of the transducer and contact FUJIFILM SonoSite or your local representative...
  • Page 19: Cleaning The Iviz Carry Case

    If the iViz ultrasound system shows damage or deterioration, such as corrosion, discoloration, pitting, or cracking, do not dispose of it. Contact FUJIFILM SonoSite or your local representative. The battery inside this unit may explode if exposed to very high temperatures. Do not destroy this unit by incinerating or burning.
  • Page 20: Compatible Cleaners And Disinfectants (System And Transducers)

    Compatible cleaners and disinfectants (system and transducers) Table 2-5: Cleaners and disinfectants compatible with the iViz system and transducers Disinfectants Airkem A-33 Alcohol-Ethanol Aniosyme DD1- 5% Anioxide 1000 Asepti Wipe II Baxedin Pre-Op Caviwipes Cidalkan Cidalkan Lingettes Clinell Universal Sanitising Wipes Control III Coverage Germicidal Surface...
  • Page 21 Table 2-5: Cleaners and disinfectants compatible with the iViz system and transducers Disinfectants Dispatch Wipes Dynacide PA Echo Clean lingettes End Bac II Endosporine Envirocide Gluco Chlor Wipes Hydrogen Peroxide 3% Incidin Plus 1% Incidin Plus 3% Korsolex Extra Medzyme MetriSponge Percept RTU Wipes Protex Spray...
  • Page 22 Table 2-5: Cleaners and disinfectants compatible with the iViz system and transducers Disinfectants Sani-Cloth Bleach Wipes Sani-Cloth HB Sani-Cloth Plus Staphene Steranios 2% Super Sanicloth Trigene Advance Wipes Tristel Solo Tristel Wipes T-Spray T-Spray II Virex TB Wet Task Wipes Wex-Cide Wipes Plus Disinfecting Wipes Compatible cleaners and disinfectants (system and transducers)
  • Page 23: Compatible Cleaners And Disinfectants (Accessories)

    Compatible cleaners and disinfectants (accessories) Table 2-6: Cleaners and disinfectants compatible with iViz accessories Disinfectants Sani-Cloth AF3 Sanicloth Bleach Wipes Sani-Cloth HB Sani-Cloth Plus Super Sanicloth Tristel Wipes T-Spray II Compatible high-level disinfectants compatible (transducers) Table 2-7: High-level disinfectants compatible with iViz transducers High-level Disinfectants Cidex Cidex Plus...
  • Page 24 Compatible high-level disinfectants compatible (transducers)
  • Page 25 Compatible high-level disinfectants (transducers) ................... 38 This chapter contains instructions for cleaning and disinfecting the X-Porte ultrasound system, transducers, and accessories. Use the FUJIFILM SonoSite recommendations when cleaning or disinfecting your ultrasound system, transducer, and accessories. Use the cleaning recommendations in the component manufacturer’s instructions when cleaning or disinfecting your components.
  • Page 26: Before Getting Started

    (PPE), such as protective eyewear and gloves.  Inspect the system to determine that it is free of any unacceptable deterioration, such as corrosion, discoloration, pitting, or cracked seals. If damage is evident, discontinue use, and contact FUJIFILM SonoSite or your local representative. ...
  • Page 27: Cleaning And Disinfecting

    Cleaning and disinfecting Determining the required cleaning and disinfecting level The cleaning instructions contained in this chapter are based on requirements mandated by the American Food and Drug Administration (FDA). Failure to follow WARNINGS these instructions may result in cross contamination and patient infection. The level of cleaning and disinfecting required for the system is dictated by the type of tissue it contacts during use.
  • Page 28: Clean And Disinfect System And Transducer To A High Level (Semi-Critical Uses)

    This can damage the transducer and void the warranty.  Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
  • Page 29 5 Clean the exterior surfaces of the ULTRASOUND SYSTEM to remove any debris or bodily fluids. Use the following procedure: a Use either a premoistened wipe or a soft cloth dampened with cleaner or disinfectant. Choose a cleaner from Table 3-2, “Cleaners and disinfectants compatible with X-Porte system and transducers” page 30.
  • Page 30 Do not immerse the transducer connector in any disinfectant solution. Cautions  Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage or discolor the transducer and void the warranty.
  • Page 31: Clean And Disinfect System And Transducer To A Low Level (Non-Critical Uses)

    12 Dispose of the disinfectant according to the manufacturer’s guidelines. 13 Examine the transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, discontinue use of the transducer, and contact FUJIFILM SonoSite or your local representative.
  • Page 32 Doing so may cause solution to leak into the system, damaging it and voiding the warranty.  Use only FUJIFILM SonoSite-recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can Cautions damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
  • Page 33: Storing The Transducer

    Allow the ultrasound system, including the console and transducer, to air dry in a clean, well-ventilated space. Examine the system, transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, do not use the transducer. Instead, contact FUJIFILM SonoSite or your local representative. Storing the transducer To store the transducer 1 Make sure the transducer has been cleaned and disinfected as detailed in this chapter.
  • Page 34: Transporting The Transducer

    Store the transducer in an environment that is safe and has good airflow. Do not store the transducer in closed containers or where condensation may occur. Avoid direct sunlight and exposure to x-rays. Recommended storage temperature range is between 0 F) and +45 C (113 If using a wall-mounted rack for storage, ensure that:...
  • Page 35 WARNINGS this chapter or according to special instructions received from FUJIFILM SonoSite. If you are returning the transducer to FUJIFILM SonoSite, document the disinfection on a “Declaration of Cleanliness, ” and attach it to the packing list.
  • Page 36: Compatible Cleaners And Disinfectants (System And Transducers)

    Compatible cleaners and disinfectants (system and transducers) Table 3-2: Cleaners and disinfectants compatible with X-Porte system and transducers Disinfectants Airkem A-33 Alcohol-Ethanol Anios Wipes Aniosyme DD1- 5% Anioxide 1000 Asepti HB Asepti Wipe II Baxedin Pre-Op Carpe Diem TB Wipes Caviwipes Cidalkan Lingettes Clinell Universal...
  • Page 37 Table 3-2: Cleaners and disinfectants compatible with X-Porte system and transducers Disinfectants Control III Coverage Germicidal Surface Wipes Coverage Plus Germicidal Surface Wipes Coverage Spray Coverage Spray TB Dynacide PA Echo Clean lingettes End Bac II Endosporine Envirocide Gluco Chlor Wipes Hydrogen Peroxide Incidin Plus 1% Incidin Plus 3%...
  • Page 38 Table 3-2: Cleaners and disinfectants compatible with X-Porte system and transducers Disinfectants MetriSponge Metrizyme Mikrobac Forte 2% Mikrobac Tissues Oxivir TB Wipes Percept RTU Protex Spray Protex Wipes Rely+On Perasafe Sani-Cloth AF Sani-Cloth AF3 Sani-Cloth Bleach Wipes Sani-Cloth HB Sani-Cloth Plus Simple Chlor Staphene Compatible cleaners and...
  • Page 39 Table 3-2: Cleaners and disinfectants compatible with X-Porte system and transducers Disinfectants Steranios 2% Steranios 20% Trigene Advance Wipes Tristel Solo T-Spray T-Spray II Virex II 256 Virex TB Virox 5 RTU Wipes Wet Task Wipes Wex-Cide Wipes Plus Disinfecting Wipes Wood Wyant Ultra Wipes Compatible cleaners and disinfectants (system and transducers)
  • Page 40: Compatible Cleaners And Disinfectants (System Accessories)

    Compatible cleaners and disinfectants (system accessories) Table 3-3: Cleaners and disinfectants compatible with X-Porte system accessories Disinfectants Accel TB Wipes Airkem A-33 Alcohol-Ethanol Alcohol - Isopropyl Anios Wipes Aniosyme DD1- 5% Anioxide 1000 Asepti HB Asepti Steryl Asepti Wipe II Baxedin Pre-Op Carpe Diem TB Wipes Caviwipes...
  • Page 41 Table 3-3: Cleaners and disinfectants compatible with X-Porte system accessories Disinfectants Cidalkan Lingettes Clinell Universal Sanitising Wipes Clorox Germicidal Wipes Clorox Hydrogen Peroxide Wipes Control III Coverage Germicidal Surface Wipes Coverage Plus Germicidal Surface Wipes Coverage Spray Coverage Spray TB DisCide Ultra Disinfecting Towlettes Dynacide PA...
  • Page 42 Table 3-3: Cleaners and disinfectants compatible with X-Porte system accessories Disinfectants Envirocide Gluco Chlor Wipes Hydrogen Peroxide 3% Incidin Plus 1% Incidin Plus 3% Medzyme Mikrobac Forte 2% Mikrobac Tissues Mikrozid Tucher NU - Cidex Oxivir TB Wipes Percept RTU Wipes Protex Spray Protex Wipes Rely+On Perasafe...
  • Page 43 Table 3-3: Cleaners and disinfectants compatible with X-Porte system accessories Disinfectants Sani-Cloth AF Sani-Cloth AF3 Sani-Cloth Bleach Wipes Sani-Cloth HB Sani-Cloth Plus Simple Chlor Staphene Steranios 2% Steranios 20% Super Sani-Cloth Trigene Advance Wipes Tristel Solo Tristel Wipes T-Spray T-Spray II Compatible cleaners and disinfectants (system accessories)
  • Page 44: Compatible High-Level Disinfectants (Transducers)

    Table 3-3: Cleaners and disinfectants compatible with X-Porte system accessories Disinfectants Virex II 256 Virex TB Virox 5 RTU Wipes Wet Task Wipes Wex-Cide Wipes Plus Disinfecting Wipes Wood Wyant Ultra Wipes Compatible high-level disinfectants (transducers Table 3-4: High-level disinfectants compatible with X-Porte transducers High-level Disinfectants Cidex Cidex OPA...
  • Page 45 Table 3-4: High-level disinfectants compatible with X-Porte transducers High-level Disinfectants Madacide-1 Compatible high-level disinfectants (transducers)
  • Page 46 Compatible high-level disinfectants (transducers)
  • Page 47 Compatible high-level disinfectants (transducers) ................... 61 This chapter contains instructions for cleaning and disinfecting the Edge ultrasound system, transducers, and accessories. Use the FUJIFILM SonoSite recommendations when cleaning or disinfecting your ultrasound system, transducer, and accessories. Use the cleaning recommendations in the component manufacturer’s instructions when cleaning or disinfecting your components.
  • Page 48 Ensure that cleaning and disinfecting solutions and wipes are not expired. WARNINGS Some cleaners and disinfectants can cause an allergic reaction to some individuals.
  • Page 49: Before Getting Started

    (PPE), such as protective eyewear and gloves.  Inspect the system to determine that it is free of any unacceptable deterioration, such as corrosion, discoloration, pitting, or cracked seals. If damage is evident, discontinue use, and contact FUJIFILM SonoSite or your local representative. ...
  • Page 50: Cleaning And Disinfecting

    Cleaning and disinfecting Determining the required cleaning and disinfecting level The cleaning instructions contained in this chapter are based on requirements mandated by the American Food and Drug Administration (FDA). Failure to follow WARNINGS these instructions may result in cross contamination and patient infection. The level of cleaning and disinfecting required for the system is dictated by the type of tissue it contacts during use.
  • Page 51: Clean And Disinfect System And Transducer To A High Level (Semi-Critical Uses)

    This can damage the transducer and void the warranty.  Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
  • Page 52 5 Clean the exterior surfaces of the ULTRASOUND SYSTEM to remove any debris or bodily fluids. Use the following procedure: a Use either a premoistened wipe or a soft cloth dampened with cleaner or disinfectant. Choose a cleaner from Table 4-2, “Cleaners and disinfectants compatible with Edge system and transducers”on page 53.
  • Page 53 Do not immerse the transducer connector in any disinfectant solution. Cautions  Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage or discolor the transducer and void the warranty.
  • Page 54: Clean And Disinfect System And Transducer To A Low Level (Non-Critical Uses)

    12 Dispose of the disinfectant according to the manufacturer’s guidelines. 13 Examine the transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, discontinue use of the transducer, and contact FUJIFILM SonoSite or your local representative.
  • Page 55 Doing so may cause solution to leak into the system, damaging it and voiding the warranty.  Use only FUJIFILM SonoSite-recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can Cautions damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
  • Page 56: Storing The Transducer

    Allow the ultrasound system, including the console and transducer, to air dry in a clean, well-ventilated space. Examine the system, transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, do not use the transducer. Instead, contact FUJIFILM SonoSite or your local representative. Storing the transducer To store the transducer 1 Make sure the transducer has been cleaned and disinfected as detailed in this chapter.
  • Page 57: Transporting The Transducer

    Store the transducer in an environment that is safe and has good airflow. Do not store the transducer in closed containers or where condensation may occur. Avoid direct sunlight and exposure to x-rays. Recommended storage temperature range is between 0 F) and +45 C (113 If using a wall-mounted rack for storage, ensure that:...
  • Page 58 WARNINGS this chapter or according to special instructions received from FUJIFILM SonoSite. If you are returning the transducer to FUJIFILM SonoSite, document the disinfection on a “Declaration of Cleanliness, ” and attach it to the packing list.
  • Page 59: Compatible Cleaners And Disinfectants (System And Transducers)

    Compatible cleaners and disinfectants (system and transducers Table 4-2: Cleaners and disinfectants compatible with Edge system and transducers Disinfectants Accel Intervention Wipes Accel TB Wipes Airkem A-33 Alcohol-Ethanol Alcohol - Isopropyl Anios Wipes Aniosyme DD1- 5% Anioxide 1000 Asepti HB Asepti Wipe II Baxedin Pre-Op Caviwipes...
  • Page 60 Table 4-2: Cleaners and disinfectants compatible with Edge system and transducers Disinfectants Coverage Germicidal Surface Wipes Coverage Plus Germicidal Surface Wipes Coverage Spray Coverage Spray TB Dynacide PA Echo Clean lingettes End Bac II Endosporine Envirocide Gluco Chlor Wipes Hydrogen Peroxide Incidin Plus 1% Incidin Plus 3% Medzyme...
  • Page 61 Table 4-2: Cleaners and disinfectants compatible with Edge system and transducers Disinfectants Mikrobac Forte 2% Mikrobac Tissues Oxivir TB Wipes Percept RTU Wipes Protex Spray Protex Wipes Rely+On Perasafe Sani-Cloth AF Sani-Cloth AF3 Sani-Cloth Bleach Wipes Sani-Cloth HB Sani-Cloth Plus Staphene Steranios 2% Steranios 20%...
  • Page 62: Compatible Cleaners And Disinfectants (System Accessories)

    Table 4-2: Cleaners and disinfectants compatible with Edge system and transducers Disinfectants Tristel Wipes T-Spray T-Spray II Virex II 256 Virex TB Virox 5 RTU Wipes Wet Task Wipes Wex-Cide Wipes Plus Disinfecting Wipes Compatible cleaners and disinfectants (system accessories) Table 4-3: Cleaners and disinfectants compatible with Edge system accessories Disinfectants Accel Intervention Wipes...
  • Page 63 Table 4-3: Cleaners and disinfectants compatible with Edge system accessories Disinfectants Airkem A-33 Alcohol-Ethanol Alcohol - Isopropyl Anios Wipes Aniosyme DD1- 5% Anioxide 1000 Asepti HB Asepti Steryl Asepti Wipe II Baxedin Pre-Op Carpe Diem TB Wipes Caviwipes Cidalkan Cidalkan Lingettes Clinell Universal Sanitising Wipes Clorox Germicidal Wipes...
  • Page 64 Table 4-3: Cleaners and disinfectants compatible with Edge system accessories Disinfectants Clorox Hydrogen Peroxide Wipes Control III Coverage Germicidal Surface Wipes Coverage Plus Germicidal Surface Wipes Coverage Spray Coverage Spray TB DisCide Ultra Disinfecting Towlettes Dynacide PA Echo Clean lingettes End Bac II Endosporine Envirocide...
  • Page 65 Table 4-3: Cleaners and disinfectants compatible with Edge system accessories Disinfectants Medzyme Mikrobac Forte 2% Mikrobac Tissues Mikrozid Tucher NU - Cidex Oxivir TB Wipes Percept RTU Wipes Protex Spray Protex Wipes Rely+On Perasafe Sani-Cloth AF Sani-Cloth AF3 Sani-Cloth Bleach Wipes Sani-Cloth HB Sani-Cloth Plus Simple Chlor...
  • Page 66 Table 4-3: Cleaners and disinfectants compatible with Edge system accessories Disinfectants Steranios 2% Steranios 20% Super Sani-Cloth Trigene Advance Wipes Tristel Solo Tristel Wipes T-Spray T-Spray II Virex II 256 Virex TB Virox 5 RTU Wipes Wet Task Wipes Wex-Cide Wipes Plus Disinfecting Wipes Wood Wyant Ultra Wipes Compatible cleaners and disinfectants...
  • Page 67: Compatible High-Level Disinfectants (Transducers)

    Compatible high-level disinfectants (transducers Table 4-4: High-level disinfectants compatible with Edge transducers High-level Disinfectants Cidex Cidex OPA Metricide OPA Plus Madacide-1 Revital-Ox Resert HLD Cidex Plus Compatible high-level disinfectants (transducers)
  • Page 68 Compatible high-level disinfectants (transducers)
  • Page 69 Also included is a listing of disinfectants that FUJIFILM SonoSite has tested and deemed compatible with the SonoSite Edge II system, transducers, and accessories. This information is presented in the following tables. ...
  • Page 70: Before Getting Started

    (PPE), such as protective eyewear and gloves.  Inspect the system to determine that it is free of any unacceptable deterioration, such as corrosion, discoloration, pitting, or cracked seals. If damage is evident, discontinue use, and contact FUJIFILM SonoSite or your local representative. ...
  • Page 71: Cleaning And Disinfecting

    Cleaning and disinfecting Determining the required cleaning and disinfecting level The cleaning instructions contained in this chapter are based on requirements mandated by the American Food and Drug Administration (FDA). Failure to follow WARNING these instructions may result in cross contamination and patient infection. The level of cleaning and disinfecting required for the system is dictated by the type of tissue it contacts during use.
  • Page 72: Clean And Disinfect System And Transducer To A High Level (Semi-Critical Uses)

    This can damage the transducer and void the warranty.  Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
  • Page 73 5 Clean the exterior surfaces of the ULTRASOUND SYSTEM to remove any debris or bodily fluids. Use the following procedure: a Use either a premoistened wipe or a soft cloth dampened with cleaner or disinfectant. Choose a cleaner from Table 5-2, “Cleaners and disinfectants compatible with SonoSite Edge II system and transducers”on page 74.
  • Page 74 Do not immerse the transducer connector in any disinfectant solution. Cautions  Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage or discolor the transducer and void the warranty.
  • Page 75: Clean And Disinfect System And Transducer To A Low Level (Non-Critical Uses)

    12 Dispose of the disinfectant according to the manufacturer’s guidelines. 13 Examine the transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, discontinue use of the transducer, and contact FUJIFILM SonoSite or your local representative.
  • Page 76 Doing so may cause solution to leak into the system, damaging it and voiding the warranty.  Use only FUJIFILM SonoSite-recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can Cautions damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
  • Page 77: Storing The Transducer

    Allow the ultrasound system, including the console and transducer, to air dry in a clean, well-ventilated space. Examine the system, transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, do not use the transducer. Instead, contact FUJIFILM SonoSite or your local representative. Storing the transducer To store the transducer 1 Make sure the transducer has been cleaned and disinfected as detailed in this chapter.
  • Page 78: Transporting The Transducer

    Store the transducer in an environment that is safe and has good airflow. Do not store the transducer in closed containers or where condensation may occur. Avoid direct sunlight and exposure to x-rays. Recommended storage temperature range is between 0 F) and +45 C (113 If using a wall-mounted rack for storage, ensure that:...
  • Page 79 WARNING this chapter or according to special instructions received from FUJIFILM SonoSite. If you are returning the transducer to FUJIFILM SonoSite, document the disinfection on a “Declaration of Cleanliness, ” and attach it to the packing list.
  • Page 80: Compatible Cleaners And Disinfectants (System And Transducers)

    Compatible cleaners and disinfectants (system and transducers) Table 5-2: Cleaners and disinfectants compatible with SonoSite Edge II system and transducers Disinfectants Accel Intervention Wipes Accel TB Wipes Airkem A-33 Alcohol-Ethanol Alcohol - Isopropyl Anios Wipes Aniosyme DD1- 5% Anioxide 1000 Asepti HB Asepti Steryl Asepti Wipe II...
  • Page 81 Table 5-2: Cleaners and disinfectants compatible with SonoSite Edge II system and transducers Disinfectants Clorox Germicidal Wipes Clorox Hydrogen Peroxide Wipes Control III Coverage Germicidal Surface Wipes Coverage Plus Germicidal Surface Wipes Coverage Spray Coverage Spray TB DisCide Ultra Disinfecting Towlettes Dynacide PA Echo Clean lingettes End Bac II...
  • Page 82 Table 5-2: Cleaners and disinfectants compatible with SonoSite Edge II system and transducers Disinfectants Incidin Plus 3% Medzyme MetriSponge Mikrobac Forte 2% Mikrobac Tissues Oxivir TB Wipes Percept RTU Wipes Protex Spray Protex Wipes Rely+On Perasafe Sani-Cloth AF Sani-Cloth AF3 Sani-Cloth Bleach Wipes Sani-Cloth HB...
  • Page 83 Table 5-2: Cleaners and disinfectants compatible with SonoSite Edge II system and transducers Disinfectants Trigene Advance Wipes Tristel Solo Tristel Wipes T-Spray T-Spray II Virex II 256 Virex TB Virox 5 RTU Wipes Wet Task Wipes Wex-Cide Wipes Plus Disinfecting Wipes Wood Wyant Ultra Wipes Compatible cleaners and disinfectants (system and transducers)
  • Page 84: Compatible Cleaners And Disinfectants (System Accessories)

    Compatible cleaners and disinfectants (system accessories) Table 5-3: Cleaners and disinfectants compatible with SonoSite Edge II system accessories Disinfectants Accel Intervention Wipes Accel TB Wipes Airkem A-33 Alcohol-Ethanol Alcohol - Isopropyl Anios Wipes Aniosyme DD1- 5% Anioxide 1000 Anioxy-Twin Asepti HB Asepti Steryl Asepti Wipe II Baxedin Pre-Op...
  • Page 85 Table 5-3: Cleaners and disinfectants compatible with SonoSite Edge II system accessories Disinfectants Control III Coverage Spray Coverage Spray TB DisCide Ultra Disinfecting Towlettes Dispatch Wipes Dynacide PA Echo Clean lingettes End Bac II Endosporine Envirocide Gluco Chlor Wipes Hydrogen Peroxide 3% Incidin Plus 1% Incidin Plus 3% Medzyme...
  • Page 86 Table 5-3: Cleaners and disinfectants compatible with SonoSite Edge II system accessories Disinfectants Rely+On Perasafe Sani-Cloth AF Sani-Cloth AF3 Sani-Cloth Bleach Wipes Sani-Cloth HB Sani-Cloth Plus Staphene Steranios 2% Steranios 20% Super Sani-Cloth Trigene Advance Wipes Tristel Solo Tristel Wipes T-Spray T-Spray II Virex II 256...
  • Page 87: Compatible High-Level Disinfectants (Transducers)

    Compatible high-level disinfectants (transducers) Table 5-4: High-level disinfectants compatible with SonoSite Edge II transducers High-level Disinfectants Cidex Cidex OPA Metricide OPA Plus Madacide-1 Cidex Plus Compatible high-level disinfectants (transducers)
  • Page 88 Compatible high-level disinfectants (transducers)
  • Page 89 Compatible high-level disinfectants (transducers) ................. 103 This chapter contains instructions for cleaning and disinfecting the S Series ultrasound system, transducers and accessories. Use the FUJIFILM SonoSite recommendations when cleaning or disinfecting your ultrasound system, transducer, and accessories. Use the cleaning recommendations in the component manufacturer’s instructions when cleaning or disinfecting your components.
  • Page 90 Ensure that cleaning and disinfecting solutions and wipes are not expired WARNINGS Some cleaners and disinfectants can cause an allergic reaction to some individuals.
  • Page 91: Before Getting Started

    (PPE), such as protective eyewear and gloves.  Inspect the system to determine that it is free of any unacceptable deterioration, such as corrosion, discoloration, pitting, or cracked seals. If damage is evident, discontinue use, and contact FUJIFILM SonoSite or your local representative. ...
  • Page 92: Cleaning And Disinfecting

    Cleaning and disinfecting Determining the required cleaning and disinfecting level The cleaning instructions contained in this chapter are based on requirements mandated by the American Food and Drug Administration (FDA). Failure to follow WARNINGS these instructions may result in cross contamination and patient infection. The level of cleaning and disinfecting required for the system is dictated by the type of tissue it contacts during use.
  • Page 93: Clean And Disinfect System And Transducer To A High Level (Semi-Critical Uses)

    This can damage the transducer and void the warranty.  Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
  • Page 94 5 Clean the exterior surfaces of the ULTRASOUND SYSTEM to remove any debris or bodily fluids. Use the following procedure: a Use either a premoistened wipe or a soft cloth dampened with cleaner or disinfectant. Choose a cleaner from Table 6-2, “Cleaners and disinfectants compatible with S Series system and transducers” page 95.
  • Page 95 Do not immerse the transducer connector in any disinfectant solution. Cautions  Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage or discolor the transducer and void the warranty.
  • Page 96: Clean And Disinfect System And Transducer To A Low Level (Non-Critical Uses)

    12 Dispose of the disinfectant according to the manufacturer’s guidelines. 13 Examine the transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, discontinue use of the transducer, and contact FUJIFILM SonoSite or your local representative.
  • Page 97 Doing so may cause solution to leak into the system, damaging it and voiding the warranty.  Use only FUJIFILM SonoSite-recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can Cautions damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
  • Page 98: Storing The Transducer

    Allow the ultrasound system, including the console and transducer, to air dry in a clean, well-ventilated space. Examine the system, transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, do not use the transducer. Instead, contact FUJIFILM SonoSite or your local representative. Storing the transducer To store the transducer 1 Make sure the transducer has been cleaned and disinfected as detailed in this chapter.
  • Page 99: Transporting The Transducer

    Store the transducer in an environment that is safe and has good airflow. Do not store the transducer in closed containers or where condensation may occur. Avoid direct sunlight and exposure to x-rays. Recommended storage temperature range is between 0 F) and +45 C (113 If using a wall-mounted rack for storage, ensure that:...
  • Page 100 WARNINGS this chapter or according to special instructions received from FUJIFILM SonoSite. If you are returning the transducer to FUJIFILM SonoSite, document the disinfection on a “Declaration of Cleanliness, ” and attach it to the packing list.
  • Page 101: Compatible Cleaners And Disinfectants (System And Transducers)

    Compatible cleaners and disinfectants (system and transducers) Table 6-2: Cleaners and disinfectants compatible with S Series system and transducers Disinfectants Accel TB Wipes Airkem A-33 Alcohol-Ethanol Alcohol - Isopropyl Anios Wipes Aniosyme DD1- 5% Asepti HB Asepti Wipe II Bacoban WB Baxedin Pre-Op Carpe Diem TB Wipes...
  • Page 102 Table 6-2: Cleaners and disinfectants compatible with S Series system and transducers Disinfectants Clorox Germicidal Wipes Clorox Hydrogen Peroxide Wipes Control III Coverage Germicidal Surface Wipes Coverage Plus Germicidal Surface Wipes Coverage Spray Coverage Spray TB DisCide Ultra Disinfecting Towlettes Dynacide PA Echo Clean lingettes...
  • Page 103 Table 6-2: Cleaners and disinfectants compatible with S Series system and transducers Disinfectants Hydrogen Peroxide Incidin Plus 1% Incidin Plus 3% Medzyme MetriSponge Mikrobac Forte 2% Mikrobac Tissues Oxivir TB Wipes Percept RTU Wipes Protex Spray Protex Wipes Sani-Cloth AF Sani-Cloth AF3 Sani-Cloth Bleach Wipes...
  • Page 104 Table 6-2: Cleaners and disinfectants compatible with S Series system and transducers Disinfectants Simple Chlor Staphene Steranios 2% Steranios 20% Trigene Advance Wipes Tristel Solo Tristel Wipes T-Spray T-Spray II Virex II 256 Virex TB Virox 5 RTU Wipes Wet Task Wipes Wex-Cide Wipes Plus Disinfecting Wipes...
  • Page 105: Compatible Cleaners And Disinfectants (System Accessories)

    Compatible cleaners and disinfectants (system accessories) Table 6-3: Cleaners and disinfectants compatible with S Series system accessories Disinfectants Accel Intervention Wipes Accel TB Wipes Airkem A-33 Alcohol-Ethanol Alcohol - Isopropyl Anios Wipes Aniosyme DD1- 5% Anioxide 1000 Asepti HB Asepti Steryl Asepti Wipe II Baxedin Pre-Op Carpe Diem TB Wipes...
  • Page 106 Table 6-3: Cleaners and disinfectants compatible with S Series system accessories Disinfectants Clorox Germicidal Wipes Clorox Hydrogen Peroxide Wipes Control III Coverage Germicidal Surface Wipes Coverage Plus Germicidal Surface Wipes Coverage Spray Coverage Spray TB DisCide Ultra Disinfecting Towlettes Dynacide PA Echo Clean lingettes End Bac II Endosporine...
  • Page 107 Table 6-3: Cleaners and disinfectants compatible with S Series system accessories Disinfectants Incidin Plus 3% Medzyme Metrizyme Mikrobac Forte 2% Mikrobac Tissues NU - Cidex Oxivir TB Wipes Percept RTU Wipes Protex Spray Protex Wipes Rely+On Perasafe Sani-Cloth AF Sani-Cloth AF3 Sani-Cloth Bleach Wipes Sani-Cloth HB Sani-Cloth Plus...
  • Page 108 Table 6-3: Cleaners and disinfectants compatible with S Series system accessories Disinfectants Staphene Steranios 2% Steranios 20% Super Sani-Cloth Trigene Advance Wipes Tristel Solo Tristel Wipes T-Spray T-Spray II Virex II 256 Virex TB Virox 5 RTU Wipes Wet Task Wipes Wex-Cide Wipes Plus Disinfecting Wipes Wood Wyant Ultra Wipes...
  • Page 109: Compatible High-Level Disinfectants (Transducers)

    Compatible high-level disinfectants (transducers) Table 6-4: High-level disinfectants compatible with S Series transducers High-level Disinfectants Cidex Cidex OPA Metricide OPA Plus Madacide-1 Revital-Ox Resert Cidex Plus Compatible high-level disinfectants (transducers)
  • Page 110 Compatible high-level disinfectants (transducers)
  • Page 111 Compatible high-level disinfectants (transducers) ................. 124 This chapter contains instructions for cleaning and disinfecting the M-Turbo ultrasound system, transducers, and accessories. Use the FUJIFILM SonoSite recommendations when cleaning or disinfecting your ultrasound system, transducer, and accessories. Use the cleaning recommendations in the component manufacturer’s instructions when cleaning or disinfecting your components.
  • Page 112: Before Getting Started

    (PPE), such as protective eyewear and gloves.  Inspect the system to determine that it is free of any unacceptable deterioration, such as corrosion, discoloration, pitting, or cracked seals. If damage is evident, discontinue use, and contact FUJIFILM SonoSite or your local representative. ...
  • Page 113: Cleaning And Disinfecting

    Cleaning and disinfecting Determining the required cleaning and disinfecting level The cleaning instructions contained in this chapter are based on requirements mandated by the American Food and Drug Administration (FDA). Failure to follow WARNINGS these instructions may result in cross contamination and patient infection. The level of cleaning and disinfecting required for the system is dictated by the type of tissue it contacts during use.
  • Page 114: Clean And Disinfect System And Transducer To A High Level (Semi-Critical Uses)

    This can damage the transducer and void the warranty.  Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
  • Page 115 5 Clean the exterior surfaces of the ULTRASOUND SYSTEM to remove any debris or bodily fluids. Use the following procedure: a Use either a premoistened wipe or a soft cloth dampened with cleaner or disinfectant. Choose a cleaner from Table 7-2, “Cleaners and disinfectants compatible with M-Turbo system and transducers”on page 116.
  • Page 116 Do not immerse the transducer connector in any disinfectant solution. Cautions  Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage or discolor the transducer and void the warranty.
  • Page 117: Clean And Disinfect System And Transducer To A Low Level (Non-Critical Uses)

    12 Dispose of the disinfectant according to the manufacturer’s guidelines. 13 Examine the transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, discontinue use of the transducer, and contact FUJIFILM SonoSite or your local representative.
  • Page 118 Doing so may cause solution to leak into the system, damaging it and voiding the warranty.  Use only FUJIFILM SonoSite-recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can Cautions damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
  • Page 119: Storing The Transducer

    Allow the ultrasound system, including the console and transducer, to air dry in a clean, well-ventilated space. Examine the system, transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, do not use the transducer. Instead, contact FUJIFILM SonoSite or your local representative. Storing the transducer To store the transducer 1 Make sure the transducer has been cleaned and disinfected as detailed in this chapter.
  • Page 120: Transporting The Transducer

    Store the transducer in an environment that is safe and has good airflow. Do not store the transducer in closed containers or where condensation may occur. Avoid direct sunlight and exposure to x-rays. Recommended storage temperature range is between 0 F) and +45 C (113 If using a wall-mounted rack for storage, ensure that:...
  • Page 121 WARNINGS this chapter or according to special instructions received from FUJIFILM SonoSite. If you are returning the transducer to FUJIFILM SonoSite, document the disinfection on a “Declaration of Cleanliness, ” and attach it to the packing list.
  • Page 122: Compatible Cleaners And Disinfectants (System And Transducers)

    Compatible cleaners and disinfectants (system and transducers) Table 7-2: Cleaners and disinfectants compatible with M-Turbo system and transducers Disinfectants Accel Intervention Wipes Accel Prevention Airkem A-33 Alcohol-Ethanol Alcohol - Isopropyl Anios Wipes Aniosyme DD1- Anioxide 1000 Asepti HB Asepti Wipe II Baxedin Pre-Op Carpe Diem TB Wipes...
  • Page 123 Table 7-2: Cleaners and disinfectants compatible with M-Turbo system and transducers Disinfectants Clorox Germicidal Wipes Clorox Hydrogen Peroxide Wipes Control III Coverage Germicidal Surface Wipes Coverage Plus Germicidal Surface Wipes Coverage Spray Coverage Spray TB Dynacide PA Echo Clean lingettes End Bac II Endosporine Envirocide...
  • Page 124 Table 7-2: Cleaners and disinfectants compatible with M-Turbo system and transducers Disinfectants Incidin Plus 1% Incidin Plus 3% Medzyme MetriSponge Mikrobac Forte 2% Mikrobac Tissues Oxivir TB Wipes Percept RTU Wipes Protex Spray Protex Wipes Sani-Cloth AF Sani-Cloth AF3 Sani-Cloth Bleach Wipes Sani-Cloth HB Sani-Cloth Plus...
  • Page 125 Table 7-2: Cleaners and disinfectants compatible with M-Turbo system and transducers Disinfectants Steranios 2% Steranios 20% Trigene Advance Wipes Tristel Solo Tristel Wipes T-Spray T-Spray II Virex II 256 Virex TB Virox 5 RTU Wipes Wet Task Wipes Wex-Cide Wipes Plus Disinfecting Wipes Compatible cleaners and disinfectants (system and transducers)
  • Page 126: Compatible Cleaners And Disinfectants (System Accessories)

    Compatible cleaners and disinfectants (system accessories Table 7-3: Cleaners and disinfectants compatible with M-Turbo system accessories Disinfectants Accel Intervention Wipes Accel TB Wipes Airkem A-33 Alcohol-Ethanol Alcohol - Isopropyl Anios Wipes Aniosyme DD1- 5% Anioxide 1000 Asepti HB Asepti Steryl Asepti Wipe II Baxedin Pre-Op Carpe Diem TB Wipes...
  • Page 127 Table 7-3: Cleaners and disinfectants compatible with M-Turbo system accessories Disinfectants Clinell Universal Sanitising Wipes Clorox Germicidal Wipes Clorox Hydrogen Peroxide Wipes Control III Coverage Germicidal Surface Wipes Coverage Plus Germicidal Surface Wipes Coverage Spray Coverage Spray TB DisCide Ultra Disinfecting Towlettes Dynacide PA Echo Clean lingettes...
  • Page 128 Table 7-3: Cleaners and disinfectants compatible with M-Turbo system accessories Disinfectants Incidin Plus 1% Incidin Plus 3% Medzyme Mikrobac Forte 2% Mikrobac Tissues NU - Cidex Oxivir TB Wipes Percept RTU Wipes Protex Spray Protex Wipes Rely+On Perasafe Sani-Cloth AF Sani-Cloth AF3 Sani-Cloth Bleach Wipes Sani-Cloth HB...
  • Page 129 Table 7-3: Cleaners and disinfectants compatible with M-Turbo system accessories Disinfectants Staphene Steranios 2% Steranios 20% Super Sani-Cloth Trigene Advance Wipes Tristel Solo Tristel Wipes T-Spray T-Spray II Virex II 256 Virex TB Virox 5 RTU Wipes Wet Task Wipes Wex-Cide Wipes Plus Disinfecting Wipes Wood Wyant Ultra Wipes...
  • Page 130: Compatible High-Level Disinfectants (Transducers)

    Compatible high-level disinfectants (transducers) Table 7-4: High-level disinfectants compatible with M-Turbo transducers High-level Disinfectants Cidex Cidex OPA Metricide OPA Plus Madacide-1 Revital-Ox Resert HLD Cidex Plus Compatible high-level disinfectants (transducers)
  • Page 131 Compatible high-level disinfectants (transducers) ................. 142 This chapter contains instructions for cleaning and disinfecting the NanoMaxx ultrasound system, transducers, and accessories. Use the FUJIFILM SonoSite recommendations when cleaning or disincomponentfecting your ultrasound system, transducer, and accessories. Use the cleaning recommendations in the component manufacturer’s instructions when cleaning or disinfecting your components.
  • Page 132: Before Getting Started

    (PPE), such as protective eyewear and gloves.  Inspect the system to determine that it is free of any unacceptable deterioration, such as corrosion, discoloration, pitting, or cracked seals. If damage is evident, discontinue use, and contact FUJIFILM SonoSite or your local representative. ...
  • Page 133: Cleaning And Disinfecting

    Cleaning and disinfecting Determining the required cleaning and disinfecting level The cleaning instructions contained in this chapter are based on requirements mandated by the American Food and Drug Administration (FDA). Failure to follow WARNING these instructions may result in cross contamination and patient infection. The level of cleaning and disinfecting required for the system is dictated by the type of tissue it contacts during use.
  • Page 134: Clean And Disinfect System And Transducer To A High Level (Semi-Critical Uses)

    This can damage the transducer and void the warranty.  Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
  • Page 135 5 Clean the exterior surfaces of the ULTRASOUND CONSOLE to remove any debris or bodily fluids. Use the following procedure: a Use either a premoistened wipe or a soft cloth dampened with cleaner or disinfectant. Choose a cleaner from Table 8-2, “Cleaners and disinfectants compatible with NanoMaxx system and transducers”on page 136.
  • Page 136 Do not immerse the transducer connector in any disinfectant solution. Cautions  Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can damage or discolor the transducer and void the warranty.
  • Page 137: Clean And Disinfect System And Transducer To A Low Level (Non-Critical Uses)

    12 Dispose of the disinfectant according to the manufacturer’s guidelines. 13 Examine the transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, discontinue use of the transducer, and contact FUJIFILM SonoSite or your local representative.
  • Page 138 Doing so may cause solution to leak into the system, damaging it and voiding the warranty.  Use only FUJIFILM SonoSite-recommended cleaners and disinfectants. Using a non- recommended disinfecting solution or incorrect solution strength can Cautions damage the System and transducer and void the warranty. Follow the disinfectant manufacturer’s recommendations for solutions strengths.
  • Page 139: Storing The Transducer

    Allow the ultrasound system, including the console and transducer, to air dry in a clean, well-ventilated space. Examine the system, transducer and cable for damage, such as cracks or splitting where fluid can enter. If damage is evident, do not use the transducer. Instead, contact FUJIFILM SonoSite or your local representative. Storing the transducer To store the transducer 1 Make sure the transducer has been cleaned and disinfected as detailed in this chapter.
  • Page 140: Transporting The Transducer

    Store the transducer in an environment that is safe and has good airflow. Do not store the transducer in closed containers or where condensation may occur. Avoid direct sunlight and exposure to x-rays. Recommended storage temperature range is between 0 F) and +45 C (113 If using a wall-mounted rack for storage, ensure that:...
  • Page 141 WARNING this chapter or according to special instructions received from FUJIFILM SonoSite. If you are returning the transducer to FUJIFILM SonoSite, document the disinfection on a “Declaration of Cleanliness, ” and attach it to the packing list.
  • Page 142: Compatible Cleaners And Disinfectants (System And Transducers)

    Compatible cleaners and disinfectants (system and transducers Table 8-2: Cleaners and disinfectants compatible with NanoMaxx system and transducers Disinfectants Accel TB Wipes Airkem A-33 Alcohol-Ethanol Alcohol - Isopropyl Anios Wipes Aniosyme DD1- 5% Anioxide 1000 Anioxy-Twin Asepti HB Baxedin Pre-Op Carpe Diem TB Wipes Caviwipes Cidalkan Lingettes...
  • Page 143 Table 8-2: Cleaners and disinfectants compatible with NanoMaxx system and transducers Disinfectants Coverage Plus Germicidal Surface Wipes Coverage Spray Coverage Spray TB Dynacide PA Echo Clean lingettes Endosporine Envirocide Gluco Chlor Wipes Hydrogen Peroxide 3% Incidin Plus 1% Incidin Plus 3% Medzyme MetriSponge Mikrobac Forte 2%...
  • Page 144 Table 8-2: Cleaners and disinfectants compatible with NanoMaxx system and transducers Disinfectants Rely+On Perasafe Sani-Cloth AF Sani-Cloth AF3 Sani-Cloth Bleach Wipes Sani-Cloth HB Sani-Cloth Plus Simple Chlor Steranios 2% Steranios 20% Trigene Advance Wipes Tristel Solo Tristel Wipes T-Spray T-Spray II Virex II 256 Virex TB Virox 5 RTU Wipes...
  • Page 145: Compatible Cleaners And Disinfectants (System Accessories)

    Table 8-2: Cleaners and disinfectants compatible with NanoMaxx system and transducers Disinfectants Wipes Plus Disinfecting Wipes Wood Wyant Ultra Wipes Compatible cleaners and disinfectants (system accessories Table 8-3: Cleaners and disinfectants compatible with NanoMaxx system accessories Disinfectants Accel TB Wipes Airkem A-33 Alcohol - Isopropyl Anios Wipes...
  • Page 146 Table 8-3: Cleaners and disinfectants compatible with NanoMaxx system accessories Disinfectants Caviwipes Cidalkan Cidalkan Lingettes Clinell Universal Sanitising Wipes Clorox Germicidal Wipes Clorox Hydrogen Peroxide Wipes Control III Coverage Germicidal Surface Wipes Coverage Plus Germicidal Surface Wipes Coverage Spray Coverage Spray TB DisCide Ultra Disinfecting Towlettes Dynacide PA...
  • Page 147 Table 8-3: Cleaners and disinfectants compatible with NanoMaxx system accessories Disinfectants Incidin Plus 1% Incidin Plus 3% Medzyme Mikrobac Forte 2% Mikrobac Tissues Mikrozid Tucher NU - Cidex Oxivir TB Wipes Percept RTU Wipes Protex Spray Protex Wipes Rely+On Perasafe Sani-Cloth AF Sani-Cloth AF3 Sani-Cloth Bleach Wipes...
  • Page 148: Compatible High-Level Disinfectants (Transducers)

    Table 8-3: Cleaners and disinfectants compatible with NanoMaxx system accessories Disinfectants Super Sani-Cloth Trigene Advance Wipes Tristel Solo Tristel Wipes T-Spray T-Spray II Virex II 256 Virex TB Virox 5 RTU Wipes Wet Task Wipes Wex-Cide Wipes Plus Disinfecting Wipes Wood Wyant Ultra Wipes Compatible high-level disinfectants (transducers Table 8-4: High-level disinfectants compatible with NanoMaxx transducers...
  • Page 149 Table 8-4: High-level disinfectants compatible with NanoMaxx transducers High-level Disinfectants Cidex Plus Compatible high-level disinfectants (transducers)
  • Page 150 Compatible high-level disinfectants (transducers)
  • Page 152 P22043-01 *P22043-01*...