Preventing Electrical Shock - FujiFilm PB2020-M Operation Manual

Endoscopic ultrasonic probe
Table of Contents

Advertisement

Chapter 1 Precautions

1.3.2 Preventing Electrical Shock

12
After cleaning, rinse off any remaining detergent solution with clean water.
If detergent solution remains, the chemical disinfection to be performed at
the next step will lose its effectiveness.
After immersing this product in disinfectant solution, remove air bubbles
completely. If any air bubbles remain, effective disinfection cannot be
achieved.
When immersing this product in disinfectant solution, immerse the entire
unit. If not properly immersed, effective chemical disinfection cannot be
achieved.
When this product is immersed in the chemical solution, remove air
bubbles on the surface. Not doing so may pose an infection risk.
Proceed with gas sterilization after vaporizing water out of this product.
If gas sterilization is performed while moisture remains in this product,
sterilization of wet parts will be incomplete.
When performing ethylene oxide gas sterilization of this product,
ensure that the surfaces are dry before attempting ethylene oxide gas
sterilization. If any moisture remains in this product, sterilization of undried
parts will be incomplete and it could be a source of infection.
When performing ethylene oxide gas sterilization of this product, detach
the connector cap from the probe connector. If sterilization is performed
while the connector cap is placed on the probe connector, sterilization will
be incomplete and may pose an infection risk.
Connect the power plug to the protective earth receptacle. Not doing so
may cause an electric shock.
This product is not intended to be used with an electrosurgical unit. If
used with an electrosurgical unit, it may cause burn injury to the patient or
end-user.
WARNING
WARNING
202B1263042