Chapter 3 Inspection Before Use - FujiFilm PB2020-M Operation Manual

Endoscopic ultrasonic probe
Table of Contents

Advertisement

Chapter 3 Inspection Before Use

Chapter 3 Inspection Before Use
WARNING
Be sure to wear sterile gloves when inspecting this product. Not doing so
may contaminate this product and pose an infection risk.
In clinical procedures in which this product that has been sterilized is
used, exercise caution not to contaminate this product when placing it
temporarily. Not doing so may pose an infection risk.
CAUTION
Do not use the equipment that shows any signs of abnormality or
irregularity. Use of abnormal equipment may lead to misdiagnosis or
increase risks to patient safety.
CAUTION
A liquid ultrasound transmission medium is sealed in this product.
Depending on the storage conditions, air bubbles may appear inside the
ultrasound head or an ultrasound transmission medium is not contained
in the ultrasound head. Check this product for air bubbles inside the
ultrasound head before use. If the ultrasound transmission medium is not
in the ultrasound head or air bubbles are present in the ultrasound head,
remove air bubbles. Clear ultrasound images cannot be obtained for air
bubbles.
23
202B1263042

Advertisement

Table of Contents
loading

Table of Contents